Maintenance & Care 
 

Upload Firmware 

When  you turn on the device using 
the  power  switch,  the  version   
number  will  appear  briefly  on  the      
display: 

 

 

1.  Download  the  firmware  from 

http://www.hadewe.net/video-library

Save  the  file  on  a  USB  stick.  The 
file  must  be  located  in  the  root   
directory.  The  file  name 

encrypt-

ed_blto_m7.bin

 

must 

not 

be 

changed. 

 

2.  Turn  on  the  device  with  the  
power  switch  and  insert  the  USB 
stick. 

 

 

3.  Simultaneously  press  the  +  /  - 
buttons  and  hold  down  while  you 
switch  on  the  device  using  the  
power switch. 

 

4. Display during the update: 

 

This  process  takes  approximately 
15 minutes. 

Never  turn  off  the  device 
during  the  update.  Other-
wise it could be damaged.

 

5.  As  soon  as  the  last  line  "Please 
restart  device"  is  displayed,  the 
update  is  complete  and  you  can 
switch  the  device  off  using  the  
power switch. 

 

6.  The  device  can  now  be  used 
again as normal. 

 

Maintenance Overview

 

What has to be done? 

How often?  
Done by 
whom? 

cleaning the chuck 

monthly by user 

 

changing the filter bag 

each 50 hours of 
operation by 
user (check after 
each 10 hours) 

 

changing the control 
unit filter 

each 200 hours 
of operation by 
user 
 

maintenance by repair 
workshop (changing 
worn parts; housing; 
cleaning; check leak-
age current) 

each 500 hours 
of operation by 
repair workshop 

Summary of Contents for Medeas

Page 1: ...Art 0980 0981 FB 04 288 Rev 2 Date 23 01 2019 Manual Medeas ...

Page 2: ...edeas Plus 17 Clamping the Tool Bur Medeas Chrome 17 Adjusting the Speed 17 Memory Keys 18 Foot Switch 18 Switsch on the Handpiece Medeas Plus 18 Select Menu 19 Help Menu 19 Settings Menu 20 Confirm Filter Change 20 Turn Light ON OFF Medeas Plus 20 Pictures Videos 21 Messages 24 Maintenance Care 25 Cleaning the Chuck Medeas Plus 25 Cleaning the Clamping System Medeas Chrome 26 Upload Firmware 27 M...

Page 3: ... and nails The handpiece is compatible with rotating tools such as diamond polishers steel cutters and ceramic cutters These can be worked with up to a speed of 30 000 rpm Medeas Chrome respectively 40 000 rpm Medeas Plus thus providing an efficient way of removing tissue The direction of rotation can be changed by pressing a button The integrated filtered suction function effectively sucks up any...

Page 4: ...e if they have been given supervision or instruction concerning use of the appliance in a safe way and under stand the hazards involved Children should be supervised not to play with the appliance Cleaning and user maintenance shall not be made by children without supervision To avoid the risk of electrocution only connect the device to the main power supply If you pull the power plug it will stop...

Page 5: ...center Prevent your hair from wrapping itself around moving parts If applicable wear a hair net When working with materials which might cre ate dust or moisture use a drilling unit with suc tion or spray technique Wear a nose and mouth protection Always follow the safety instructions provided by the manufacturer of the materials being formed and used For safety reasons you must always wear dis pos...

Page 6: ...equipment assure that the tool can be pushed far enough into the chuck test this with a dipstick and clean if neces sary the inserted tool sits properly in the chuck and will not cant jam while clamping it the tool is approved for using it at the max speed see manufacturer s instructions or only use the speed required by the instru ment the tool is absolutely dry For safety reasons do not change t...

Page 7: ... particles of liquids Always follow the operating instructions Do not carry out any repairs modifications or mainte nance work yourself This is only to be accom plished by an authorized specialist The mains cable may also only be changed by the manu facturer or an authorized specialist Mainte nance records may be supplied to repair work shops and trained technicians upon request Never remove the h...

Page 8: ...t will run with the rotation speed and suc tion power used last Take care to not start the mill at an overly high rotation speed Select the required rotation speed before changing the milling tool In case the touch display shows a damage or does not work properly the device may not be used any longer Please send it to a qualified service workshop for repair ...

Page 9: ...After every treatment please carry out the following steps Remove the rotating tool e g bur capper etc first Then handle the tool according to the instructions of its manufac turer Remove tissue remains by wiping off the unit with a mildly moistened paper cloth Afterwards clean all surfaces with dis infection cloth e g Mikrozid AF Tücher of Schülke and Mayr There is no limit to the number of times...

Page 10: ...essed switch unit off standby 5 speed display 6 time 7 date 8 changing the direction of rotation left right switch 9 maintenance required 10 USB stick plugged in green light 11 Medeas Plus light yellow when switched on 12 13 14 memory keys for speed suction 15 suction stage set 16 bringing up menu with additional functions 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 ...

Page 11: ...or the following scopes foot care milling trimming and polishing manicure grinding milling and polishing works The device is determined of operating rotating instru ments which will not penetrate the skin or murcous membrane if used appropriate but could evoke damages accidently Caution In the area of pedicure you have to work with suction techniques Operation The working area must conform to the ...

Page 12: ... display prompts you to set the device language This is followed by 5 other pages with important clues With con tinue you proceed to the next page On the last page press to start the device The counter for calculat ing the maintenance proposal will automatically be set to zero The display will shut down now and the unit is in standby Take the handpiece in hand If the hose is twisting the handpiece...

Page 13: ...esulting milling dust is swirled in a clockwise direction in the field of view Now switch to suction By means of the icon 15 you can adjust the strength of suction Frequently used speed and suction capacity can be saved First set the desired speed and suction Save this by pressing one of the three memory keys 12 13 or 14 for 4 seconds To recall the stored speed and suction power quickly briefly pr...

Page 14: ...14 Usage Work Positioning The following images are examples of how to work on various parts of the foot Always make sure that the foot is firmly gripped and that you can work freely and without strain ...

Page 15: ...15 Usage ...

Page 16: ...hours Ex change a full filter bag immediately A filter bag should be exchanged at least after 50 operating hours At first unfold the paper of the filter bag art 5179 and push your finger into the opening in the front Pull out the lid of the dust chamber and remove the old filter Hold the filter in the middle to prevent it from bending and sliding off Therefore you avoid it from coming off again It...

Page 17: ...he tool if the handpiece is turned off Clamping the Tool Bur Medeas Chrome Clamp the tools bur only when the handpiece is not working Insert the tool bur until it stops Ones can use all tools burs which has a shaft according to DIN EN ISO 1797 1 with a diameter of 2 35 mm For security reasons only change the tool if the handpiece is turned off Adjusting the Speed Take the handpiece with tools in t...

Page 18: ...er can be switched ON OFF Briefly press the left memory key 12 to recall the speed suction Another quick press of the central lock key 13 will recall the speed suction Another quick press of the first memory key 12 can also be used This only work with tools that are approved for the 12 to 14 stored speeds When the display is switched off you can turn it on by keeping the foot switched pressed keep...

Page 19: ...for setting the time date and language By means of this button you can return to the previous menu Help Menu Manual Here is a short form of the manual This does not replace the original manual in paper form as the latter contains more detailed instructions and safety instructions Insert Burs this clip demonstrates the clamping of rotary tools Clean Chuck video clip that demonstrates the cleaning o...

Page 20: ...tivated Confirm Filter Change Every 50 hours of operation you will get the message to Check filters After the filter change the following must be entered into the instrument On the start screen click the icon for the select menu Then click Change Filter Confirm the filter change with YES Turn Light ON OFF Medeas Plus To turn the light on or off click the top of the bar on the light icon Alternativ...

Page 21: ...maximum of 100 files may be located in the directory for the image viewer Click Picture Video in the select menu to start the viewer File list Scrollbar Thumbnail Reset In the file list you can see the files from the USB stick By scrolling down you can browse through the list Select an image and a small preview is shown on the right Click on thumbnail to enlarge it Click once more on the enlarged ...

Page 22: ...eo library we provide a program ready to convert your own videos from other formats to EMF format Open the file to begin installation Follow the instructions on the monitor The conversion program to be used is File Converter Input Select a source file The following formats are supported MP4 WMV The source files should be proportionate to 4 3 to be taken so as to avoid distortion when converting Ou...

Page 23: ...ive in order to consider the devices In the video library you can find videos and pictures that help you to explain the treatment to your customers Especially new customers are often unse cure because they do not know what to expect during a treatment It is helpful to explain the treatment steps to the patient This is often difficult especially when the patient cannot see the treatment site Pictur...

Page 24: ...Tapping this icon will display the following message In this case contact your distributor or repair shop REMINDER CHECK FILTER Every 50 hours of operation in the suction mode you will receive the message Please change fil ter You will find instructions in the chapter Exchange Micro Filter and a help video under the menu item Help SHUTDOWN DUE TO SPEED DIFFERENCE The device switches off when the s...

Page 25: ...sh For opening the chuck push the button forward Set the screw spanner wrench on the axle and hold it tight Now use the front of the second spanner to unscrew the chuck To assemble the chuck follow the steps in reverse order Make sure that the chuck is opened during the assembly the handpiece button must be in forward position and is firmly applied afterwards How to clean the chuck Soak the chuck ...

Page 26: ...tem This may cause that instruments can no longer be inserted deep enough and that a tight grip is no longer granted The result of this often is abrasive wear and therefore the clamping system is to be cleaned every 50 hours of operation For this purpose screw off the pro tective sleeve of the handpiece Clean the inside of protective sleeve and the hull with a dry scrubber or brush Now insert care...

Page 27: ...ay during the update This process takes approximately 15 minutes Never turn off the device during the update Other wise it could be damaged 5 As soon as the last line Please restart device is displayed the update is complete and you can switch the device off using the power switch 6 The device can now be used again as normal Maintenance Overview What has to be done How often Done by whom cleaning ...

Page 28: ...s of the handpiece in form of rust bloom and causes pitting Therefore when you not working with the handpiece always remove the tool bur from it Always make sure that you do not let the handpiece fall down or exposes it to any shocks The ballbearings and the motor could be damaged by this Only work with tools which are flawless Do not work with tools which are bent or out of balance whose shaft is...

Page 29: ...oved easily Chuck is dirty Clean the chuck according to the instruc tions in the manual Rubber under the knob is tilted Turn rubber about 90 Unit does not work at all Check if the main switch of the unit is turned on at the side of the unit Footswitch is used constantly or faulty Pull out the socket of the footswitch to see if this is the cause Pull out the mains plug to reset the device The unit ...

Page 30: ...ure 500hPa to 1060hPa 0981 Medeas Chrome control unit total weight 2 7kg dimension W273xH142xD186mm input 100V 240V AC voltage 50 60Hz power consumpt 1A 100V 0 5A 240V fuse primary F1 F2 4A delay action fuse fuse Sec Thermal fuse protection class I accuracy of speed display 10 handpiece weight 140g diameter 19 25 mm length 118 mm speed 6000 30000 rpm Accessories Spare Parts art 0970 footswitch 375...

Page 31: ...ing the disassembling of the device Also wear a protective mask for your face according EN 14683 version IIR or EN 149 FFP2 Disposal of filter The filter bag and the control unit filter are considered low contami nated waste Dispose those in intransparent liquid tight resistant plastics sacks and do not dispose them compressed in the regular household garbage Symbols Caution Important note Follow ...

Page 32: ...hadewe GmbH Grambartstraße 10 30165 Hannover Germany info hadewe de www hadewe de ...

Reviews: