Operating Instructions
9.1
Measurement
1. Point the meter at the object to be measured.
2.
Hold down the trigger to begin measurement. The “
” symbol
flashes on the top of the display to indicate measurement is
taking place.
3. The main measurement display show the current temperature
reading of the measured surface.
4. Release the trigg
er and “HOLD” symbol will appear in the upper
left corner and the display will hold the last reading.
5.
The meter’s auto off function will turn off the meter 30s after
releasing the trigger.
Note:
- Environmental temperature can affect the accuracy of the reading.
It is recommended the meter be placed in the measurement
environment for 30 min. prior to taking measurement.
- Laser can be turned off for measurements at close range to
conserve battery power.
9.2
Settings
Emissivity: default emissivity is 0.95
To change emissivity: press “MODE” until the bottom left shows the “E”
symbol with two arrows pointing up and down. Press the up and down
buttons to adjust the emissivity. Emissivity range: 0.1-1.0.
To show the maximum recorded value: press “MODE” until “MAX”
appears on the bottom left corner of the display.
To show the minimum recorded value: press “MODE” until “MIN” appears
on the bottom left corner of the display.