background image

Step 8: Sanitize, Test and Purge

SANITIZE

  (

Note: Sanitization is recommended immediately after RO Filter System installation and any 

inner-part servicing. The person sanitizing should have clean hands during this process.)

1.  Shut off water supply to RO system.
2.  Open faucet. If tank is not empty, allow 

to drain until empty.

3.  With included eyedropper and house-

hold bleach (5.25%), disconnect white 
tubing from the manifold outlet labeled 
“TANK”. Note: Bleach needs to be 
handled according to manufacturer’s 
instructions.

4.  Add 3ml bleach into open end of tank 

white tubing.

5.  Reconnect the tank from the tube into 

the manifold outlet labeled “TANK”. 
Be sure to push tubing in all the way.

6.  Sanitation will be completed during the following pressure test and purge. Important: 

Bleach must be completely removed from system before drinking water. See Purge 
instructions below.

PRESSURE TEST

  (

Important: Complete sanitization prior to pressure test.)

1.  Open cold water supply valve to RO Filter System.
2.  To purge air from the plumbing system, open kitchen faucet. Close faucet when water 

runs smooth.

3.  Confirm RO faucet is closed.
4.  Within approximately 2 hours, pressure will start to build in the RO Filter System. Care-

fully inspect all connections and fittings while this pressure buildup occurs.

5.  Check for leaks. If leaks are found, fix by ensuring all tubing is cut squarely and fully 

inserted. Also confirm there are no scratches, dents or notches at tubing end. If there 
are, squarely cut 1” off and re-insert.

 

 Note: When RO Filter System is first pressurized, water may project from faucet air gap hole until air is 

passed from RO Filter System.

PURGE

1.  Open RO faucet and let water flow through system for 24 hours.

 

 Note: Flow rate will be slow during this time.

2.  Close RO faucet after purge is complete. 

 

 Note: Your RO Filter System is ready for use when purge is complete, however, you will not have filtered 

water immediately. It takes 1-3 hours to completely fill the tank. The flow rate will be less than your 

kitchen faucet. Water will run to the drain while the RO Filter System is filtering water – even when not in 

use. This is normal. Water going to drain will stop

 

Summary of Contents for AquaH20 HF-RO

Page 1: ...AquaH20 Drinking Water System Owner s Manual MODEL HF RO...

Page 2: ...out G RO Faucet Body Stem stem not shown H Water Storage Tank I Eye Dropper J Tank Connector K Teflon Tape L Drain Connector M Nuts Bolts N Foam Seal O O Ring Faucet Base P Toggle Bolts Q Flow Restric...

Page 3: ...o ensure adequate space and proper positioning 3 8 white tubing will be used to connect manifold to tank and from faucet stem to remineralizer Measure out 3 8 white tubing from the top of the tank to...

Page 4: ...e tube in place tighten the compression nut to compress the plastic sleeve and create a seal NOTE Use a wrench to ensure complete seal Avoid over tightening 5 Do not connect the other end at this time...

Page 5: ...e tubing through the nut collar and sleeve into the connector B Unscrew the compression nut from the tank connector to ensure that the tubing is connected all the way through the collar and sleeve C S...

Page 6: ...k hole until faucet base is flush with countertop Position toggle bolts under sink surface being certain faucet stem will not be obstructed Loosely fasten faucet base to countertop by tightening toggl...

Page 7: ...ocation 2 Remove protective cover from back of foam seal 3 Knock center hole out align holes and attach to front plate of drain connector 4 Allowing room for drilling position the drain connector on s...

Page 8: ...ing in order to prevent crimping kinks loops or folds TUBING DONT S DO NOT cut tubing too short Always double check measurements before cutting DO NOT bend or crimp or kink tubing DO NOT discard exces...

Page 9: ...lled in step 4 from the faucet into the remineralizer port labeled OUTLET IV Manifold to Tank 3 8 white tubing already attached to tank 1 Take the white tubing leading from the storage tank and insert...

Page 10: ...ttach red carbon sump to the 1st stage position on the inlet side of the system manifold Make sure all con nection points are aligned and push the top of the sump up and into the system manifold Turn...

Page 11: ...System 2 To purge air from the plumbing system open kitchen faucet Close faucet when water runs smooth 3 Confirm RO faucet is closed 4 Within approximately 2 hours pressure will start to build in the...

Page 12: ...conditions on the above are 250 ppm TDS and a 30 6 recovery rate Production rate and TDS reduction figures are for a new Membrane that has been rinsed for 24 hours The production rate of a new Membran...

Page 13: ...rapidly convert trivalent arsenic to pentavalent arsenic Other water treatment chemicals such as ozone and potassium permanganate will also change trivalent arsenic to pentavalent arsenic A combined c...

Page 14: ...igher than 8 0 membrane life may be as short as 12 months If pH is under 7 5 membrane life may last up to 36 months When output water quality and production rate decrease it is time to replace the fil...

Page 15: ...It is recommended the restrictor assembly be periodically inspected to be sure it is clean and unrestricted If service is required on the drain flow assembly disassemble and reassemble as outlined in...

Page 16: ...ternational against NSF ANSI Standards 42 and 53 in model HF RO for the reduction claims specified on the Performance Data Sheet as verified and substantiated by test data and at nsf org The HF RO sys...

Reviews: