9
•This product should be inspected by a dedicated personnel every day and record its operation
condition. If the temperature inside the refrigerator is out of acceptable range, transfer the
goods to other refrigerators until the fault is eliminated.
•This product is a medical refrigerator and the set temperature is 5 °C. Please confirm that the
stored goods match the environment inside the refrigerator to avoid damage to the goods and
economic loss.
•Due to the refrigeration inertia, the displayed temperature of this product may differ from the
actual temperature inside the refrigerator, which is normal.
•Please do not place any articles between the refrigerator floor and the bottom shelf of the unit,
so as to avoid blocking the air duct and affecting the uniformity of the temperature inside the
refrigerator.
•Refrigerators are designed for storing products. Do not load in too much warm product at one
time. Overloading the refrigerator will reduce the system’s expected life. Therefore, product
and samples should be placed into the refrigerator in batches.
•Do not damage the refrigeration circuit.
•Do not use unlicensed appliances inside the refrigerator.
The refrigerator has memory of the set values. In case of power on after power failure, the refrigerator
will continue to operate according to the set parameters before the power failure.
•Once the refrigerator is powered off, wait for at least five minutes before restart it, so
as to avoid damage to the compressor or system.
•When the refrigerator is not in use for a long time, remove the power plug and turn off
the battery switch to prevent electric shock, electric leakage or fire due to aging of the
plug cord.
•If the refrigerator is left unused for a long time in an unsupervised area, make sure
that no child can get access to it and the door is not completely closed.
Operation after recovery from power failure
Warning