16
P6
USB derivative mode
(0 is all, 1 is the last 1
month, 2 is the last 2
months, and so on)
0
12
1/Month
0
SS
User management
level parameter setting
Press “Set” button, 000 flickers and it requires to enter
the registration / user management lever password of
the IC card. Enter the correct password and then press
“Set” to enter the user management level parameters.
If you enter the wrong password for 3 times in a row,
it will exit and redisplay the temperature inside the
refrigerator.
The followings are user management level parameters
IC
IC card registration /
user management level
password modification
000
999
1
008
CL1
IC card logout
(all IC cards)
Press the “Set” button, 000 flickers and it requires
to enter the registration password. Enter the correct
password and then press “Set” to execute the order.
The CL1 flickers for 3 times and there is a buzz sound,
which means that the logout is completed. If you enter
the wrong password for 3 times in a row, it will exit and
redisplay the temperature inside the refrigerator.
Lt
Refrigeration light
delay turn-off time
(button light)
-1 (cancel
delay turn-
off), 1
60
1/Min
8
dA
Door opening buzz
alarm delay time
1
30
1/Min
5
FA
Buzzer volume
1
3
1
3
S1/S2
Ozone sterilization
control parameter
Reserved function, not enabled
TT
Temperature unit
0 (Celsius)
1 (Fahrenheit), reserved
function, not enabled
0