• Working Temperature:0

~35

℃;

Debug and Use

Preparation before use

• According to the actual power supply situation, select the appropriate power cord, random with AC 
power cord, battery power supply cable, Car-charger power supply cable.

• Connect the main power supply.

 

Turn on the battery switch.

Power on

When  the  battery  is powered, the  built-in  battery  switch can  be  switched off  to 
extend the battery's service time.

Summary of Contents for HZY-40Z

Page 1: ......

Page 2: ...s 1 Application guidelines 4 Product Features 5 Installation 6 Components Display panel 7 Application Method 9 Debug and Use 10 Alarm 12 Clean and disinfection 13 Maintenance routine 15 FAQ 16 Circuit diagram 17 Specification 18 ...

Page 3: ...ly be operated by trained and authorized person Please use the special power supply indicated on the nameplate of the transport cooler otherwise it may cause fire or electric shock When the voltage is less than 100 V or above 242 V an automatic voltage regulator should be installed Ensure that the voltage meets the requirements If the power cord be lengthened the cut off area of the lengthened line ...

Page 4: ...t medicines or suspended particles inside and around the transport cooler are not inhaled during maintenance Otherwise there is a risk of health hazards If the transport cooler is not in use for a long time the power plug should be unplugged and the backup battery switch should be turned off to prevent electric shock leakage or fire caused by the aging of the power cord If the transport cooler is id...

Page 5: ...ssemble repair or modify the transfer box by himself Otherwise fire or injury may be caused by improper operation Do not place flammable explosive dangerous goods or volatile items in the transfer box or use flammable spray near the transfer box otherwise it may cause explosion or fire Metal objects such as nails or wire shall not be inserted into any openings and gaps in the transfer box or into any ...

Page 6: ...above material is likely to damage the painted surface with plastic rubber parts Be careful not to wipe plastic rubber parts with volatile solvents like gasoline The stainless steel parts inside the transfer box are not acid resistant please pay attention to the anti corrosion measures Do not use acid media in the box The transfer box is not used for a long time the power supply including backup b...

Page 7: ...ture Alarm sensor faults alarms Multiple Alarm type Buzzer and light Safety ground 3 Cooling system Deep Optimized Compression Mechanism Cooling System Better Refrigeration Performance 4 Humanization design humanized design microprocessor control system intelligent worry free easy to adjust 100mm thick foaming layer better insulation performance Drainage port design easy to clean the condensate an...

Page 8: ...ransport cooler is used beyond the scope of the product working environment We can t ensure that it can achieve the expected performance Installation Installation environment Space Reserved Use appropriate tools such as scissors or wallpaper knife cut the packing tape remove the screws on the outer packing box with a screwdriver then remove the packing Check the accessories Unpacking and handling ...

Page 9: ...7 Components Display panel Components Main body Castor Display panel Handle Aviation plug Door Lock catch ...

Page 10: ... or under voltage 2 Cooling Green Lighting compressor is working 3 Alarm Red Flashing Alarm occurs 4 Heating Yellow lighting Heater is working 5 Battery Green flashing Battery power is less than 20 6 Network Green Lighting Normal Flashing disconnection more than 1min Cooling Heating Power Alarm Network Battery Set Back ...

Page 11: ... will open Electric lock Under user setting mode when LED displays UC Press Set button LED displays 000 Press or 06 to input password 06 if the password is correct LED displays off Press or LED diplays on or off Press Set button to confirm your choice UCt means the working time for UVC light time range is 5 to 30mins When LED displays UCt press Set button LED displays the value Press or to adjust the...

Page 12: ...ituation select the appropriate power cord random with AC power cord battery power supply cable Car charger power supply cable Connect the main power supply Turn on the battery switch Power on When the battery is powered the built in battery switch can be switched off to extend the battery s service time ...

Page 13: ... of the first power on press set button buzzer alarm sound cancel default 30 min if the alarm conditions are not lifted the sound alarm will work again until the temperature reaches the set interval 20 10 or 2 8 or 10 20 Factory default setting value of cooler is 4 if you need to adjust please refer to the instructions to adjust accordingly After a period of running the temperature in the chamber i...

Page 14: ...r faults Control sensor E30 Buzzer and light Display sensor E31 Buzzer and light Ambient sensor failed E32 Buzzer and light Condenser sensor failed E34 Buzzer and light Defrost sensor failed E35 Buzzer and light Alarm and alarm code Alarm recovery The cooler has Alarm recovery function When the inner temperature trigger alarm value Alarm occurred Pressing Set button can cancel the buzzer it can no...

Page 15: ...y the surface thoroughly with clean cloth Cleaning the display Clean the display with 100 microfiber dry cloth Pollution Control Process During the use of the transfer box the user needs to develop the corresponding pollution control process compatible with the transfer box according to the following requirements Wipe spray disinfection standard manual disinfection control procedures for transport ...

Page 16: ... power supply before cleaning and disinfection work a Use the power switch to close the transport cooler b Unplug to prevent accidental connection c Ensure that the transport cooler is cut off Health hazards The surface of the studio may be contaminated Contact with contaminated cleaning fluid may cause infection Disinfectants may contain harmful substances When cleaning and disinfecting be sure to ...

Page 17: ...ioning and inspection to ensure high precision culture environment parameters If you also want the transfer box to work at the highest accuracy please contact Haier engineer In the course of routine maintenance and maintenance only Haier or Haier designated agencies can inspect or provide spare parts for the maintenance of transport cooler Before notifying the maintenance engineer please clean and...

Page 18: ...ll unresolved after operation please contact Haier after sales service Do not maintain and dismantle the transport cooler by yourself Faults Troubleshooting The transport cooler doesn t work Whether the input power meets the requirement Whether the plug and socket are in loose contact Whether the input line and control line break down The transport cooler doesn t cool down or warm Check the settin...

Page 19: ...17 Circuit diagram ...

Page 20: ...50 60Hz DC12 24V 90W Power Consumption kW h 24h 1 5 Alarm High low temperature alarm sensor error alarm power abnormal alarm door ajar alarm low battery alarm Built in battery for Alarm only 12V 7Ah USB Standard fitting Electromagnetic lock Optional Light 12V 1W UV Lamp UVC 12V 2 4mW 473nm 479nm Made Date Software version See barcode V1 Service life 10 years Anti shock Safety Classication I PS Tech...

Page 21: ...19 Name HZY 40Z basket 1 Power cord AC DC 1 User manual 1 Plastic bag 1 castor 4 Packing list ...

Page 22: ...20 ...

Page 23: ...21 ...