American Hakko Products, Inc.
13.
Carefully rotate the top half of the unit to re-assemble the two halves and give you a clear view
of the LCD display.
14.
Connect the test leads of the Yokogawa 2786 Decade Resistance Box or equivalent device to
the HAKKO FG-101 where the one test lead is connected to the test plate of the HAKKO FG-101
and the other test lead is connected to the grounding terminal of the HAKKO FG-101.
15.
Verify the calibration by setting the output value on the Yokogawa 2786 Decade Resistance Box
or equivalent device to the values in the table below and confirm that the display shows the
standard values corresponding to each output setting.
Output Value
(
Ω
)
Standard Value
(
Ω
)
0
0.0 to 0.1
2
1.9 to 2.1
40 39.5
to
40.5
16.
If there is any deviation outside the standard values, perform the calibration again.
17.
When finished, turn off the HAKKO FG-101. Remove the test leads of the Yokogawa 2786
Decade Resistance Box or equivalent device from the HAKKO FG-101. Re-secure the two halves
of the unit by installing the 5 Phillips screws along the front and rear of the casing.