2. Remove the bottom of the electromagnetic valve.
3. Clean the parts that have flux attached to them, such as the inside of the electromagnetic valve, with
alcohol.
Clean with alcohol.
Caution
Do not use thinner when cleaning.
4. Assembly is done in the opposite order of disassembly.
*There is one more electromagnetic valve. Remove the screws on the back of the chassis, and from that
point on remove and clean following the same steps.
Remove
the screws.
●
Cleaning the pump head
1. Remove the valve and valve guard and remove any attached flux.
Caution
・
When the valve guard is difficult to remove,
please warm it with hot air. Please do not try
to forcibly remove it with a screwdriver, etc.
If the valve guard becomes deformed, it will
no longer be airtight.
・
Please clean with either alcohol or thinner.
2. Install the valve and valve guard.
Caution
When assembling the pump, please make sure to keep it airtight so that there are no air leaks.
Valve
Valve guard
Clean the pump head, valve
and valve guard.
Pump head disassembly
*It is resting on its side.
*Please replace the valve if it is
deformed or hardened.
Pump head
7. MAINTENANCE
14
Summary of Contents for fm-206
Page 17: ......