background image

4. Operation

 (cont'd)

4-3.  Operation

CAUTION

Place the handpiece into the iron holder and then turn on the power.

The following display appears after turning the power on.

Lock icon

Temperature 

icon

Offset icon

Operation 

button

Communication 

icon

Setting  

temperature

Temperature 

Correction Mode

Temperature 

Setting Mode

Preset icon

Sleep Mode

Back  

button

Confirm 

button

Down  

button

Up  

button

Parameter 

icon

Tip sensor 

temperature

Preset No. 

Mode

 

Changing the temperature setting

Push 

Push this button once to display [ ] and transition to "temperature setting mode." 

This mode is used when changing the set temperature.

Push 

 once

[ ] and [

]  

disappear.

Temperature 

control begins

Normal screen

To change to 400ºC

Push 

 five times

Push

Push

Push

 Once the normal screen appears, push the       button to transition to preset No. mode.

 

Changing the preset No.

You can register up to five frequently used setting temperatures on the product, and then select the registration No. to change

the setting temperature.

Push 

Push this button once to transition to "preset No. mode."

Select one of the five temperatures registered in this mode.

( Factory default temperature settings:  P1  250ºC (600ºF), P2  300ºC (700ºF), P3  350ºC (750ºF), 

P4  400ºC (800ºF), P5  450ºC (850ºF))

Push 

 once

Temperature 

control begins

Normal screen

To change to P4 (400ºC)

Push

NOTE

The registered temperature of each preset No. can be changed in "parameter No. 

."

(See the separate document "Quick Parameter Settings")

NOTE

If you want to limit setting temperature changes, change the setting in "parameter No. 

."

(See the separate document "Quick Parameter Settings")

 

Tip temperature correction (offset)

Push 

Push this button once to display [

] and transition to "temperature correction mode." If the setting 

temperature and the measured value of the tip temperature differ in this mode, you can correct the temperature.

(Correction range: ± 50ºC/± 90ºF)

Push 

 five times

Temperature 

control begins [

] disappears.

Normal screen

To correct a temperature by 5ºC for a setting of 400ºC (to correct 

actual tip measurement reading 395ºC when set to 400ºC)

0 to 5

(ºF: 0 to 9)

0 to 9

Push

Push

Push

Push

–/0

*Note that temperatures that exceed the correction range cannot be entered.

Once a tip degrades from wear, the tip temperature tends to drop. 
The tip temperature changes if you replace the tip, the offset will 
need to be readjusted. Make sure to change the offset value as 

needed while monitoring the actual tip temperature.

You can automatically change the offset value using a HAKKO 
thermometer with a temperature sending function. 

 

Push the   

   button before sending the measured value. 

 

(See the figure on the right)

 

Pausing work (sleep mode)

Push 

Push this button once to immediately transition to "sleep mode (state where the tip temperature has 

dropped to the set activation temperature)." Use this function occasionally to prevent tip oxidation.

( Factory default setting: Parameter No. 

 setting is enabled and the activation temperature in 

Parameter No.

 is 200ºC)

Temperature control begins

Normal screen

Transition to normal mode

Lift the handpiece from the iron holder.

  Push any of the operation buttons.

  The product will not transition to sleep mode in the following cases:

When the setting temperature is lower than 300ºC (570ºF)

When the parameter No.   setting is disabled

  

Approximately six minutes after placing the handpiece on the iron holder, the product automatically transitions to sleep mode.

  

Change parameter No.   and 

 settings as necessary for your work. (See the separate document "Quick Parameter 

Settings")

  To further prevent tip oxidation, set auto shut-off.

(1) Enable the parameter No.   setting.

(2) 

 Set the time until the product is automatically shut-off in parameter No. 

.  

The shorter the set time, the more effective it is. 

 

If you disable the parameter No.   setting, the product will not automatically shut-off even when the set time has 

elapsed.

FX-971

Temperature sending screen

The communication icon is 

only displayed at this state.

IR receiver

5.  Parameter Settings

You can change the parameter settings to suit different work environments. See the separate document 

 

"Quick Parameter Settings."

6.  Maintenance

CAUTION

Do not file oxidation attached on the tip. This will shorten the tip lifespan.

Conducting maintenance will help keep the product in good condition and prolong the usage of the unit.

 

Daily maintenance

Setting  

temperature

Using the product at a temperature that is higher than necessary can accelerate tip deterioration 

and damage parts that are susceptible to heat. Use the lowest temperature whenever possible.

Before 

 

beginning work

Perform a visual check of the tip. Replace it if it is bent or considerably worn. Use the cleaning 

sponge to wipe off any oxidation or old solder from the tip. Impurities on a circuit board can result 

in poor soldering.

When pausing 

work

Use sleep mode instead of leaving the handpiece set to a high temperature for a long period of 

time. This prevents tip oxidation which helps to maintain workability, which can extend the tip  

lifetime. Turn off the power switch when not using the product for a long period of time.

(See "

 Pausing work (sleep mode)" in "4-3. Operation")

After finishing 

work

Thoroughly clean the tip with the cleaning sponge and then coat it with new solder. Doing so can 

prevent oxidation of the tip.

 

Periodic maintenance

Tip

Wear and tear on the tip will vary due to the operating temperature as well as the quality and amount of 

solder/flux used. Maintenance should be performed based on what suits your usage.

(1) Turn the power ON.

(2) Set the temperature to 250ºC (482ºF).

(3) Once the temperature is stable, use the cleaning sponge to wipe the tip.

(4) 

 If there is any black oxidation on the solder plating, apply new solder containing flux and then wipe it off 

with the cleaning sponge. Repeat this process until the oxidation is removed. Afterward, coat it with new 

 

solder.

(5) 

 Turn the power off and remove the tip once it has cooled. 

 

If you find flux, debris, and other particulates on anything other than the end of the tip, wipe it off with 

industrial alcohol.

Handpiece

Remove flux, debris, and other particulates adhering to 

 

the nipple.  

It may cause contact failure inside the handpiece.

Iron holder

 Press down the release button and remove the iron holder base, then clean the collected solder waste 

from the iron holder.

Rotate the cleaning wire as need to a clean side where solder is not accumulated.

FX-9701

Nipple

7.  Troubleshooting

CAUTION

Before performing an inspection or replacing parts, make sure to disconnect the power plug from the outlet.

No operation even if 

power switch is turned 

ON.

Has the power cord or connection 

plug been removed?

Plug unit into outlet.

Is the fuse blown?

Replace the fuse.

 

If the fuse is blown again, send the main unit 

(including handpiece, power cord) back for 

service.

[C - E] is displayed.

Is an incompatible soldering  

iron connected?

Connect the compatible handpiece.

Has the handpiece plug been 

removed?

Turn off the power switch, reconnect the 

handpiece, and turn the power switch back on 

again.

[H - E] is displayed.

Is the heat capacity of the tip too 

small for the object to be soldered?

Use a tip with a larger heat capacity.

Is the set value for the low temp 

alarm too small?

Increase the set value. 

( See the separate document "Parameter No.

in "Quick Parameter Settings")

[H S E] is displayed.

Is the tip an applicable genuine tip?

Turn off the power switch, insert an applicable 

genuine tip, and turn on the power switch again. 

If the problem persists, replace the tip.

[S - E] is displayed.

Is the tip fully inserted?

Insert tip firmly into the handpiece. 

 

(Do not use excessive force)

Is the heating element/sensor 

disconnected?

Measure the resistance between the heating 

element and sensor, and if the measured value is 

abnormal, replace the tip. The normal resistance 

values are as follows: 

 

T39: 5.7 

Ω

 ± 10

 (at room temperature)

 

T50: 8.0 

Ω

 ± 10

 (at room temperature)

T50

T39

Measure the resistance between these points.

Measure the resistance between these points.

[- - -] is displayed.

Is there a strong noise source 

around the soldering iron?

Move the soldering iron away from the noise 

source, or use other circuit for the power.

Cannot get solder on 

the tip.

Is the tip setting temperature too 

high or too low?

Set an appropriate temperature.

Is there any oxidation on the tip?

Remove the oxidation. (See "6. Maintenance")

The tip temperature is 

too high/low.

Is the offset value entered correct?

Measure and adjust the value.

 

(See "

 Tip temperature correction (offset)" in 

"4-3. Operation")

The auto shut-off 

 

function does not work.

Is parameter No.   [OFF]?

Change it to [On] to enable feature.

Visit the website for more support information including information for replacement parts/options. 

 

If you cannot find a solution in this manual or on the website, or if another problem occurs, 

please contact the retailer where you purchased the product.

https://www.hakko.com/doc_fx971-ah

Reviews: