Chapter 2
Installing the QE65000
Overview
You must install the OOIBase32 software application prior to connecting the QE65000 Spectrometer to
the PC. The OOIBase32 software installation installs the drivers required for QE65000 installation. If you
do not install OOIBase32 first, the system will not properly recognize the QE65000.
If you have already connected the QE65000 to the PC prior to installing OOIBase32, consult
Chapter 3:
for information on correcting a corrupt QE65000 installation.
QE65000 Installation
This section contains instructions for connecting the QE65000 in USB mode. To connect the QE65000 to
a PC via the USB port, the PC must be running the Windows 98/ME/2000/XP operating system.
►
Procedure
Follow the steps below to connect the QE65000 to a PC via the USB port:
1.
Install OOIBase32 on the destination PC.
2.
Locate the USB cable (USB-CBL-1) provided with the QE65000.
3.
Insert the square end of the cable into the side of the QE65000.
4.
Insert the rectangular end of the cable into the USB port of the PC.
5.
If you intend to run the TE Cooler, plug the +5 VDC power supply into the QE65000.
If you installed OOIBase32 prior to connecting the QE65000, the
Add New Hardware Wizard
appears
and installs the QE65000 drivers. If the drivers do not successfully install (or if you connected the
QE65000 to the PC before installing OOIBase32), consult
Chapter 3:
Configuring the QE65000 in OOIBase32
Once you install the QE65000, you must configure OOIBase32’s
Configure Spectrometer
options so
that OOIBase32 recognizes the QE65000 Spectrometer. Consult the
OOIBase32 Spectrometer Operating
Software Operating Instructions
for detailed instructions on configuring the spectrometer in OOIBase32
(see
The following sections contain instructions on initially configuring the QE65000 the first time you start
OOIBase32.
220-00000-000-02-0605
7