Chapter 4
Sample Experiments
Overview
The following sections contain information on conducting sample experiments using the QE65000 and
OOIBase32.
Preparing for Experiments
Follow the procedure below to configure the QE65000 and OOIBase32 for experiments.
►
Procedure
1.
Double-check that you have correctly installed the QE65000, installed OOIBase32, and
configured the light source and other sampling optics.
2.
Open the OOIBase32 application.
3.
Select
Spectrometer | Configure
from the menu bar, and double-check that
A/D Interface
settings are correct.
4.
Check your spectrometer setup configurations in OOIBase32:
a.
Locate the Wavelength Calibration Data sheet that came with the QE65000.
b.
Select
Spectrometer | Configure
from the menu and choose the
Wavelength Calibration
page.
c.
For each spectrometer channel in the system, enable the channel and make sure the First
Coefficient, Second Coefficient, Third Coefficient and Intercept correspond to those of the
system.
5.
Adjust the acquisition parameters using the
Acquisition Parameters
dialog bar or select
Spectrum | Configure Data Acquisition
from the menu.
If you have followed the previous steps and started OOIBase32, the spectrometer is already
acquiring data. Even with no light in the spectrometer, there should be a dynamic trace displayed
in the bottom of the graph. If you allow light into the spectrometer, the graph trace should rise
with increasing light intensity. This means the software and hardware are correctly installed.
Once you install the hardware, configure the software, and establish your sampling system, you are ready
to take measurements.