7. Routine cleaning.  

[Back to contents]

 

 

Heat exchanger fans. 

If the LiDAR is in a dusty environment, it may be necessary to blow the dust from around the heat exchanger fans with 
an air duster. The fans should also be checked routinely to make sure that they are both spinning without obstruction. 
 
 

Wiping brush.  

The wiping brush can be removed if it is full of dust, or badly deformed. There are 2 small grub screws that hold in 
place. The brush can be cleaned in water, and then set straight whilst wet and then dried. It should then maintain the 
shape. A replacement brush is available if it is damaged. 
 
 

Vinyl tape. 

The vinyl tape around the bottom of the white enclosure is not a seal – it is there just to prevent a build up of dust from 
becoming trapped between the enclosure to base plate junction. Replacement vinyl tape can be used to replace it if it 
becomes damaged. Regular tape as used for car striping is recommended 1” wide. 
 
 

Optical output window.

 

 

The optical performance will not be affected too badly from beaded standing water on the output window as illustrated 
below provided the fractional coverage of the clear aperture is not too high.  
  
 
 
 
 
 
 
 
 
 
 
 
 

 

The output window has a hard anti-reflection coating on the outer surface. The coating is reasonably robust but should 
it be deemed necessary to clean it, the procedure below should be adopted: 
 
The surface of the window should be flooded with IPA or an equivalent optical cleaning solution and left to soak for a 
short period of time to loosen any debris which might have adhered to the surface. Wiping of the window should be 
done with a soft lens tissue. This cycle should be repeated until a satisfactory result is achieved. 
 

LiDAR enclosure.

 

If the LiDAR’s enclosure becomes excessively dirty, then it can be washed down with a sponge and water. Detergents 
used for washing dishes for example, should be fine to use. Scouring pads or abrasive cleaning products should not 
be used. 
 

Purging. 

If the internal humidity of the system becomes high, then it may be necessary to purge the system with dry air. An air 
inlet on the 

connector panel

, and one of the enclosure screws will be removed during the purge.  

A suitable connector for use with the purge inlet will have been supplied with the system. If it can’t be found, this is the 
part that is needed: 

http://uk.rs-online.com/web/p/hose-couplings/3872640/

  

Please contact Halo for recommended purge procedure. 

 

 

Summary of Contents for Stream Line XR

Page 1: ...ntinuous and unattended mode of operation Typical applications include boundary layer wind profiling plume dispersion analysis of complex flows cloud studies cloud base measurements and gust and air q...

Page 2: ...irus is off Windows time syncing is disabled Daylight saving adjustment is disabled Video resolution is 1280 x 1024 BIOS settings The PC is set to come on with power The PC will come on at 12 20 PM if...

Page 3: ...to be observed at all times The LiDAR should be sited in such a way as to minimise the amount of heat generated by close surfaces such as tarmac or sunlight falling on it The temperature specification...

Page 4: ...nsit Case Packing and Unpacking 2 Connections 3 Power requirements and the UPS module 4 Powering up 5 Orientation and levelling 6 Site considerations 7 Routine cleaning 8 Typical thermal characteristi...

Page 5: ...he appropriate manual handling health and safety guidelines Unpacking reverse the process when packing The transit case consists of 3 separate pieces The LiDAR is packed fully assembled Remove the top...

Page 6: ...that it be left at this position The small pillars with rubber coned ends act as support legs when the 4 main legs are not attached They prevent the weight of the LiDAR being taken by the cooling agg...

Page 7: ...coming from the power supply module F 1 10A 5x20mm fuse for the LiDAR circuit F 2 10A 5x20mm fuse for the heaters and the thermo electric cooler TEC circuit if fitted There are 5 small LEDs close to...

Page 8: ...er un wired smaller connector and the white purging line The arrow on the power connector faces away from the LiDAR when it is the correct orientation The connector is a turn bayonet type Ensure that...

Page 9: ...e to the UPS and if a power cut happens a warning message will be displayed in the software tab area which will warn that a power cut is detected and the time remaining until the LiDAR shuts down With...

Page 10: ...e main connector body using a left hand thread Recommended PSU for connection to the 24V DC input power protection module It is recommended that the no load voltage of the PSU be turned up to 25 5V be...

Page 11: ...internal PC will be fully booted approximately 1 minute later indicated by the flashing green LED on the connector panel cluster If the unit has been allowed to cool down to below 15 C then additional...

Page 12: ...degrees Note For CSM scan files which are defined in motor encoder counts not degrees the offset bearing will not be used and all movements are with respect to the home position zero counts It will be...

Page 13: ...the heat generated underneath the LiDAR from direct sunlight a turfed area is ideal or shading if possible Ambient temperatures quoted in the temperature specification of the LiDAR are with respect to...

Page 14: ...coverage of the clear aperture is not too high The output window has a hard anti reflection coating on the outer surface The coating is reasonably robust but should it be deemed necessary to clean it...

Page 15: ...n If fuse F 2 is not blown then there may be a fault with the cooling unit in which case the LiDAR should be powered down to avoid the risk of it over heating Passive temperature option The heat excha...

Page 16: ...passwords needed to access the system are either removed or are sent to Halo prior to sending the system The system is packed in the same way that it was received The Unpacking section can be used for...

Reviews: