background image

 

P29 Instruction Manual 

 

16 

 

LC-Display (optional) 

The measured values and the menu are shown on the display. The top line is the information 
line. This line displays the type of measured value or the status of a zero-point calibration. The 
measurement value will be displayed on the central line. If the permitted measuring range is 
exceeded or not reached, the figures are replaced by corresponding arrows. The bottom line 
displays the unit, with the first two digits showing the status of the two warning messages. 
Press the Menu key 

“     “ to bring up the menu. 

In Menu mode, the parent menu item always appears in the top display line. The middle line 
always  shows  the  current  sub-menu  item  or  value  to  be  changed. The bottom  line  displays 
units or other help texts.  
Start the Menu mode by pressing the "Menu" key. It can be protected with a four-character 
password.  At  the  main menu  level,  “Menu”  is  displayed  in  the  top  line  and  “Display”  in the 
middle line. Select your desired sub-menu by pressing the 

“▲“ and “►“ keys. Press “Enter” to 

go to a sub-menu or enter a val

ue. Press the “Menu” key to go to the next menu level or to 

stop entering values. 
The following sub-menus are available in the first level: 

  Display 

  Scaling 

  Warning 

  Settings 

7.1 Display 

The display unit can be selected with this menu item. Press the 

“▲“ and “►“ keys to select the 

unit  you  want  and  press  “Enter”  to confirm. The type  of  display  (pressure, flow  rate  etc.)  is 
selected automatically. 
The following units are available: 

  Pa 

  hPa 

  kPa 

  mbar 

  mmH²O 

  mmHg 

  psi 

  inH²O 

  inHg 

  m³/s 

  m³/h 

  kg/s 

  kg/min 

  kg/h 

  m/s 

  mph 

  f/s 

  f/min 

  km/h 

7.2 Scaling 

The P29 is usually supplied with standard measuring ranges. You can use the scale to adjust 
the measuring range for your own use. The output voltage or currents are then copied to this 
scaled  range. The  scaled  range  should always  be  more 

than 0.1 times of that of the P29’s 

measuring  range,  because  otherwise  the resolution  of  the  outputs  and  the  accuracy  will  be 
poorer.  
With this menu item, there are 4 sub-menu items: 

  Pressure 

  Volumetric flow 

  Mass flow 

  Flow rate 

 Pressure 

This is an important menu item. It is used to specify the pressure range, which indicates the 
output value. This pressure range forms the basis for many other settings such as the limits 
for the warnings and factors for the flow display. 
There are two default settings for the pressure scale: 
Top = Pressure at which the output has its maximum value (e.g. 5V, 10V or 20 mA) 
Bottom = Pressure at which the output has its minimum value (e.g. -5V, 0V, 0mA or 4mA) 
 
Because the values can be freely assigned, settings such as 0V at 0Pa or 10V at -250Pa are 
also possible. 

Summary of Contents for P29

Page 1: ...Pressure Transducer Type P29 for Natural Gas halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E Mail info halstrup walcher com Internet ww...

Page 2: ...ion New 2 5 Services 3 1 flush revised New 3 3 leak test 3 3 shipping now 3 4 chapter 10 problem added All illustrations labelled New 10 1 Check fuse New 10 2 Replace fuse 9 Specifications overload ca...

Page 3: ...Instrument description 11 4 1 Features 11 4 2 User interfaces 11 4 3 Internal ports and keys 12 4 4 Front view 13 4 5 Readout of volume mass flow and flow velocity 14 Zero point calibration cycle 15 5...

Page 4: ...20 Zero point calibration 20 CS creep suppression 20 Lightning 20 Colour change 20 Brightness 20 Contrast 21 Read factory settings 21 Code 21 List of commands 22 Technical data 24 Trouble shooting Se...

Page 5: ...efully It is especially important to read and follow the notes on starting up in systems that use flammable gases Contact the manufacturer if you do not understand any part of this instruction manual...

Page 6: ...dicated on the rating plate and in the Technical data section of this manual The instrument may only be handled as indicated in this manual Modifications to the instrument are prohibited The manufactu...

Page 7: ...ent errors may occur if the instrument is not kept protected from sunlight Specific safety precautions are given in individual sections of this manual 2 5 Troubleshooting maintenance service repairs d...

Page 8: ...lead to bodily injury up to and including death if the corresponding instructions are not followed CAUTION This warns you of a potential hazard that could lead to significant property damage if corres...

Page 9: ...fore connecting the instrument to the gas lines both pressure lines should be flooded with gas otherwise the instrument itself must be flooded with gas for a longer period of time Appropriate equipmen...

Page 10: ...le pressure test For the leak test shut off the two pressure lines to the P29 and note the initial pressure display or output signal Observe the differential pressure for 1 minute The differential pre...

Page 11: ...terface Relay Keyboard Analogue output signal figure 4 1 functional description 4 2 User interfaces The following four keys are available Menu UP Right Enter Key Meaning Measurement Mode Meaning Menu...

Page 12: ...mass or volume flow Press the key to display the current value for the operating hours meter 4 3 Internal ports and keys figure 4 2 Internal connectors and push buttons Output signal Power supply figu...

Page 13: ...5V 5V 2 0V 12 0V 5 0V 7 0V 4 N 24V starts external zero point calibration Power supply No Designation Description 11 GND Ground 12 24VDC Power supply 24VDC 13 RS232 interface for PC RS232 interface f...

Page 14: ...s square rooted for example is the differential pressure value generated by a dynamic pressure probe at volume flow Analog current output signal 0 20 mA or 4 20 mA Analog voltage output signal 0 10 V...

Page 15: ...e adjusted in the Settings menu Zero point calibration can also be started with the Enter key if the device is in Measurement mode The P29 does not respond to keys being pressed during calibration 5 1...

Page 16: ...ing Warning Settings 7 1 Display The display unit can be selected with this menu item Press the and keys to select the unit you want and press Enter to confirm The type of display pressure flow rate e...

Page 17: ...le Value Unit 7 2 2 1 Value This is used to set the volumetric flow value displayed at the maximum scaled pressure 0 is always used as the lower scale value on the volumetric flow display root extract...

Page 18: ...f min km h 7 3 Warning INFORMATION For safety reasons relay outputs are not possible with the P29 However exceeding or falling below the switching thresholds can be signalled visually and acoustically...

Page 19: ...eded With this parameter you can set the duration of this warning signal for each warning The maximum value here is 1000 ms continuous tone If both warnings are active the longer warning signal of the...

Page 20: ...rameter to 0 deactivates automatic zero point calibration CS creep suppression This parameter specifies the value for creep suppression as a percentage If this value falls short of the measured pressu...

Page 21: ...ble Read factory settings The factory settings are established in the instrument before delivery With this menu item they can be reactivated if required Code You can use this menu item to specify an a...

Page 22: ...alSG Flow rate m s at max scale value ScalO Min 0 floating PRelai1 Switching pressure relay 1 Pa 120 120 of the measurement range floating RRelai1 Switching direction for Relay 1 1 decreasing 0 off 1...

Page 23: ...01 2 0 1 3 0 2 4 0 5 5 1 unsigned int UnitD Unit display 0 Pa 1 hPa 2 kPa 3 mbar 4 mmH2O 5 mmHg 6 Psi 7 inH2O 8 inHg 9 m s 10 m h 11 kg s 12 kg min 13 kg h 14 m s 15 mph 16 f s 17 f min 18 km h unsig...

Page 24: ...ssure 100 kPa for all measurement ranges Ambient conditions Medium Air non aggressive gases Operating temperature 10 C to 50 C Storage temperature 10 C to 70 C Electrical data Rated input approx 6 VA...

Page 25: ...cknowledge 10 1 Checking the fuse If the device does not work and in the case of devices with a display nothing is shown on the display the fuse may be defective To check the fuse follow the steps bel...

Page 26: ...following steps To change the fuse follow the steps below 1 Make sure that the device is no longer connected to the power supply 2 Pull the fuse upwards 3 Then reinsert the new fuse accordingly only u...

Page 27: ...P29 Instruction Manual 27 Dimension drawing figure 11 1 dimensional drawing...

Page 28: ...Pa mbar mmH2O mmHg psi inH2O inHg m s m h kg s kg min kg h m s mph f s f min km h Scaling Pressure Top xxx xx Bottom xxx xx Unit Pa hPa kPa mbar mmH2O mmHg psi inH2O inHg Vol Flow Value xxx xxxx Unit...

Page 29: ...n Italian French Output 4 20 mA 0 20 mA 5 5V 0 10 V Filter xxxxx ms key tone on off Resolution max 0 01 0 1 0 2 0 5 1 Zero point calibration xx min CS x x Lighting Off White Blue Colour change On Off...

Page 30: ...P29 Instruction Manual 30 Certificate of Conformity NEU_7100 004324_P29 docx 02 2023 Ts...

Page 31: ...P29 Instruction Manual 31 Notes...

Page 32: ...P29 Instruction Manual 32 Notes...

Reviews: