Measuring Screen 

 

 

Battery Indicator

 

The battery indicator displays the battery strength graphically. Four bars represent 

100% of charge in the battery; three bars 75%; two bars 50%; one bar 25%. No bars 

signify a low battery status and simultaneously the alarm will buzz as a warning. 

Charging of the battery is necessary at this level and after a few seconds of the warning 

sound the instrument shuts itself off automatically. 

 

Time and Date 

The current date and time is always displayed in the format of year-month-day and hour: 

minute: second, respectively. Date and time can be changed in the Setting screen. 

 

Temperature & Relative Humidity (T and R/H) 

Temperature and Relative Humidity automatically displays when an external 

temperature and humidity sensor probe is attached. 

 

• 

Use 

BACK

 keypad to toggle the concentration unit between ppm and 

mg/m

if the sampling is not started. 

• 

Push 

RUN/STOP 

keypad

 

to start continuous measuring/sampling, During 

the sampling process, the backlight will be off automatically. 

• 

Data are logged or current sampling value will be saved every time when 

the

 ENTER

 keypad is pressed. Push 

RUN/STOP 

keypad

 

to stop measuring 

and the backlight will be on again.     

  Measuring         

CO:       102 

ppm   

2008-09-06   08:40:25 

Temp: 23

°

C   RH:60% 

 

                                

Summary of Contents for HAL-HCO106

Page 1: ...1 of 1 HAL HCO106 Handheld Carbon Monoxide Meter Monitor Operational Manual 625 E Carroll Avenue Glendora CA 91741 USA Phone 510 579 8540 Fax 626 236 9246 Info haltechnologies com http haltechnologies...

Page 2: ...Page 2 of 2 This Page intentionally left as blank...

Page 3: ...ortant Messages 4 I Introduction 6 1 1 Features 7 1 2 Specifications 7 II Basic Operation 9 2 1 Measuring Screen 9 2 2 Browsing Screen 11 2 3 Setting Screen 13 2 4 Calibration 15 III Warrantee 17 Cont...

Page 4: ...rica Copyright 2008 by Hal Technology All rights reserved No part of the contents of this manual may be reproduced transmitted stored or translated into any other language in any form by any means wit...

Page 5: ...xpose to environments where rust or radioactivity are present DO NOT expose to an environment exceeding the specified limits Technical Support and Warrantee Within a year from the date purchased the m...

Page 6: ...oxide meter is a compact personal monitor that can provide a rapid indication of hazardous airborne carbon monoxide levels at simple touch of a button The HalTech HCO106 Carbon Monoxide Meter makes it...

Page 7: ...ation External digital temperature and humility sensors to assure accurate measurement Excess limit warning user defined No less than 6 hours of continuous operation 1 2 Specifications Target Gas Carb...

Page 8: ...00 240VAC to 5VDC 1A Dimension 80 W 150 H 36 D mm Weight About 200 grams Environmental Condition Operating 5 40 C 90 RH Storage 20 50 C 90 RH Standard accessories AC adapter USB cable CD with data dow...

Page 9: ...or back out of the selection The bottom of the enclosure includes USB Interface Connect to the USB interface to a computer for data downloading remote sampling or firmware upgrading Contact the manuf...

Page 10: ...month day and hour minute second respectively Date and time can be changed in the Setting screen Temperature Relative Humidity T and R H Temperature and Relative Humidity automatically displays when a...

Page 11: ...cally turn off the backlight after about several seconds if there is no keypad action The instrument will be automatically turned off if there is no keypad action after about 8 minutes 2 2 Browsing Sc...

Page 12: ...ets stored Delete the record Press an arrow keypad to move up or down to select the record to be deleted Use the arrow keypad to select Delete one or Delete all Then press ENTER keypad to delete the c...

Page 13: ...he excess exposure limit warning The user may input any value at the increment of 1ppm between 0 and 999ppm For your reference two levels of limit for carbon monoxide are in common 1 Threshold Limit V...

Page 14: ...uncertain 25 ppm Maximum allowable TWA exposure limit OSHA Used in personal CO alarms 35 ppm Maximum allowable workplace exposure limit for an 8 hour work shift NIOSH 40 hour work week 50 ppm Maximum...

Page 15: ...o be calibrated The values in the third column are calibration coefficients The example of calibration procedures is described below Here we use a zero air and the vapor concentrations of 100 ppm of c...

Page 16: ...reen Figure 8 Calibration screen 3 6 ScreenPress RUN STOP keypad to sample the correspondence value in the Y column will display the response of the concentration for current liquid standard Press RUN...

Page 17: ...normal operation due to instrument needing calibration Limitation of Warranty A Hal Technology warrants that all equipment shall be free from defects in material and workmanship under normal use for a...

Page 18: ...n by an authorized service technician or on site at buyer s facility by a Hal Technology authorized employee Buyer pays shipping costs to factory seller will pay standard return shipping costs during...

Page 19: ...ir own expense Contact HAL TECHNOLOGY LLC 625 E Carroll Avenue Glendora CA 91741 USA Phone 510 579 8540 Fax 626 963 7302 Info haltechnologies com http haltechnologies com Information Record Model __ S...

Page 20: ...l address below info haltechnologies com User Registration Form Company _____ Contact Person ______________ Address ______ ______ __ City ____State Province Country _____ Postal Code _________ Phone _...

Reviews: