WI-25023
23
PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E80]
▲
Measured Temperature Exceeds
Setpoint
Possible high impedance or
desiccated tissue at probe tip.
OR
E80 –
▲
Temp Control Error:
Check
Tmeas against Final TSET during and
after ramp (v4.1 & v4.11 only)
1. Carefully reposition and stabilize
cannula or introducer
2. Check probes, cables and
grounding pad connection
3. If problem persists, remove probe
from cannula/introducer, wipe off tip
and replace
4. Initiate new procedure and check
for proper function
5. Swap out probe, cable and
grounding pad as appropriate
6. Turn unit “OFF” then “ON” to
attempt recovery
7. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E81]
▲
Self-test failure: Temperature
Measurement
Check probe and cable connections.
Possible high impedance or
desiccated tissue at probe tip.
ERROR [E82]
▲
Excessive RF Current Measured
Possible short between electrodes,
or defective probe or cable.
1. Check power set-point
2. Check probes, cables and
grounding pad connection
3. Initiate new procedure and check
for proper function
4. Carefully reposition and stabilize
cannula or introducer
5. Swap out probe, cable and
grounding pad as appropriate
6. If problem persists, remove probe
from cannula/introducer, wipe off tip
and replace
7. Initiate new procedure and check
for proper function
8. Try new probe(s) and cable(s)
9. Turn unit “OFF” then “ON” to
attempt recovery
10. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E83]
▲
Excessive RF Voltage Measured
Check probe and cable connections.
Possible intermittent cable
connection or loss of tissue contact.
ERROR [E84]
▲
Excessive RF Power Measured
Check probe and cable connections.
Possible intermittent tissue contact
or defective probe(s) or cable(s).