PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E87]
▲
Unrecognized Probe Type
See User Manual for a list of valid
probes. Probe type may not be
enabled. Contact technical support
if problem persists
1. Check connection of adapter and
probe cables ensuring that they are
dry (ensure proper autoclave drying
time is observed)
2. See User Manual for a list of valid
probes to ensure use of the correct
probe
3. Ensure that the correct probe type
on the PMG screen is enabled
4. Try new probe(s) and cable(s)
5. Turn unit “OFF” then “ON” to
attempt recovery
6. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E88]
▲
Invalid Temperature Reading
Please check probe and cable
connections.
Possible defective probe or cable.
Try new probe and cable if problem
persists.
ERROR [E89]
▲
Temperature Limit Exceeded
Power setting may be too high. Try
using a lower power setting
1. Check power setting and adjust as
appropriate.
2. Verify temperature limit setting
(only applies to Manual Power)
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E90]
▲
Inconsistent Settings
TIME setting less than INITIAL TIME.
See ADVANCED IDL SETTINGS and
ensure INITIAL TIME is < total time
ERROR [E91]
▲
Inconsistent Settings
PEAK TEMP setting less than INITIAL
TEMP.
See ADVANCED IDL SETTINGS
and ensure INITIAL TEMP is < peak
temperature
ERROR [E92]
▲
Inconsistent Settings
TIME setting less than RAMP TIME.
See ADVANCED LESION SETTINGS
and ensure RAMP TIME is < total time