WI-25023
26
PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E93]
▲
High Impedance Detected
Check probe and cable connections.
Probe or cable may be defective.
1. Check probes, cables and
grounding pad connection
2. Initiate new procedure and check
for proper function
3. Carefully reposition and stabilize
cannula or introducer
4. Inject saline and re-insert probe
5. Initiate new procedure and check
for proper function
6. Swap out probe, cable and
grounding pad individually
7. If problem persists, remove probe
from cannula/introducer, wipe off tip
and replace
8. Initiate new procedure and check
for proper function
9. Try new probe(s) and cable(s)
10. Turn unit “OFF” then “ON” to
attempt recovery
11. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E95]
▲
Low Impedance Detected
Check probe and cable connections.
Possible short circuit in probe or
cable.
ERROR [E94]
▲
Invalid Temperature Reading
Check probe and cable connections.
Possible defective probe or cable.
Try new probe and cable if problem
persists
1. Check connection of adapter and
probe cables ensuring that they are
dry (ensure proper autoclave drying
time is observed)
2. See User Manual for a list of valid
probes to ensure use of the correct
probe
3. Ensure that the correct probe type
on the PMG screen is enabled
4. Try new probe(s) and cable(s)
5. Turn unit “OFF” then “ON” to
attempt recovery
6. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3