PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E100]
▲
Incorrect Switch Position
Examine external switch position,
and make appropriate change.
Failure detected when using TDP
placement mode
1. Check probe and cable
connections
2. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
Table of Error, Fault & Warning Codes
Code
Message
Recommended Action
(Complete List)
ERROR [E101]
▲
Pump Malfunction
1. Ensure pump lids are fully closed
2. Check pumps cable connections
3. Ensure fluid lines/tubing is
properly connected and free from
obstruction/kinks
4. Check pump cable connection
from back of pump to back of PMG
ERROR [E102]
▲
Temperature Tracking Error (>
10°C for 1 Second)
1. Carefully reposition and stabilize
cannula or introducer
2. Ensure fluid circuitry is properly
connected and free from
obstruction/kinks
3. Try new probe(s) and cable(s)
4. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E103]
▲
Dynamic Probe-Pump Mapping
Failure
1. Ensure pump lids are fully closed
2. Check pumps cable connections
3. Ensure fluid lines/tubing is
properly connected and free from
obstruction/kinks
4. Inspect for fluid drop in burette(s)
and flow from burette through the
probe(s) and back to the burette.
ERROR [E104]
▲
Pump Current Limit
Check pump unit and ensure cable
is securely connected. Contact
Technical support if problem
persists.
1. Check pump cable connection
from back of pump to back of PMG
2. Try a new pump cable
3. Replace pump with a new pump
4. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E105]
▲
No Tubing in Pump.
Check tubing for proper placement
in pump.
Check pump unit and ensure tubing
is properly install and securely
connected