WI-25023
31
PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E121]
▲
IDL Not Connected
Check probe and cable connections.
Probe or cable(s) may be defective.
1. Check probe and cable
connections
2. Try new probe(s) and cable(s)
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E122]
▲
IDL and Secondary
Thermocouple Not Connected
Check probe and cable connections.
Probe(s) or cable(s) may be
defective.
ERROR [E123]
▲
Invalid Probe A Temperature
Reading
Check probe and cable connections.
Probe or cable(s) may be defective.
Note: Only applies to Transdiscal
(TD)
1. Check connection of adapter
and probe cables ensuring that
they are dry (ensure proper
autoclave drying time is observed)
2. See User Manual for a list of
valid probes to ensure use of the
correct probe
3. Ensure that the correct probe
type on the PMG screen is enabled
4. Try new probe(s) and cable(s)
5. Turn unit “OFF” then “ON” to
attempt recovery
6. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E124]
▲
Invalid Probe B Temperature
Reading
Check probe and cable connections.
Probe or cable(s) may be defective.
Try new probe and cable(s) if
problem persists.
Note: Only applies to Transdiscal
(TD)