WI-25023
32
PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E125]
▲
High Impedance Detected
Check probe and cable
connections. Possible poor tissue
contact. Probe(s) or cable(s) may be
defective.
1. Check probes, cables and
grounding pad connection
2. Initiate new procedure and
check for proper function
3. Carefully reposition and stabilize
cannula or introducer
4. Inject saline and re-insert probe
5. Initiate new procedure and
check for proper function
6. Swap out probe, cable and
grounding pad individually
7. If problem persists, remove
probe from cannula/introducer,
wipe off tip and replace
8. Initiate new procedure and
check for proper function
9. Try new probe(s) and cable(s)
10. Turn unit “OFF” then “ON” to
attempt recovery
11. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E126]
▲
Low Impedance Detected
Possible short circuit between
electrodes, or damaged introducer
insulation.
ERROR [E127]
▲
Invalid CRP A Temperature
Reading
Check probe and cable connections.
Probe or cable(s) may be defective.
Try new probe and cable(s) if
problem persists.
1. Check probe and cable
connections
2. Try new probe(s) and cable(s)
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E128]
▲
Invalid CRP B Temperature
Reading
Check probe and cable connections.
Probe or cable(s) may be defective.
Try new probe and cable(s) if
problem persists.
ERROR [E129]
▲
Select Probe
Multiple Probes Connected. Multiple
probe connectivity not supported in
this mode. Disconnect extra probes
to continue.
1. Disconnect extra probes to
continue
2. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3