PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E150]
▲
Multi-RF TRUE-TX Ramp time
extension limit exceeded
Connect one probe at a time to
Multi-RF cable and treat each site
individually.
1. Connect one probe at a time to
Multi-RF cable and treat each site
individually
2. Carefully reposition and stabilize
cannula or introducer
3. Swap out probe(s) and cable(s)
4. If problem persists, remove probe
from cannula/introducer, wipe off tip
and replace
5. Try new probe(s) and cable(s)
6. Adjust TRUE-TX extension time
limit and ramp time in ADVANCED
SETTINGS
7. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E160]
▲
Probe Limit Exceeded on Left
Pump
Reduce number of probes
connected on left pump to TWO or
fewer. Contact technical support if
problem persists.
1. Reduce number of probes
connected on left/right pump head
to two or fewer
2. Check probe and cable
connections
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E161]
▲
Probe Limit Exceeded on Right
Pump
Reduce number of probes
connected on right pump to TWO
or fewer. Contact technical support
if problem persists.
ERROR [E162]
▲
Probe A Pump Mapping
Temperature Did Not Change by at
least 1.5°C.
1. Ensure pump lids are fully close
2. Check pumps cable connections
3. Ensure fluid lines/tubing is
properly connected and free from
obstruction/kinks
4. Inspect for fluid drop in burette(s)
and flow from burette through the
probe(s) and back to the burette.
ERROR [E163]
▲
Probe B Pump Mapping
Temperature Did Not Change by at
least 1.5°C.
1. Ensure pump lids are fully closed
2. Check pumps cable connections
3. Ensure fluid lines/tubing is
properly connected and free from
obstruction/kinks
4. Inspect for fluid drop in burette(s)
and flow from burette through the
probe(s) and back to the burette.