WI-25023
34
PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E164]
▲
Probe C Pump Mapping
Temperature Did Not Change by at
least 1.5°C.
1. Ensure pump lids are fully closed
2. Check pumps cable connections
3. Ensure fluid lines/tubing is
properly connected and free from
obstruction/kinks
4. Inspect for fluid drop in burette(s)
and flow from burette through the
probe(s) and back to the burette.
Table of Error, Fault & Warning Codes
Code
Message
Recommended Action (Complete
List)
ERROR [E165]
▲
Probe D Pump Mapping
Temperature Did Not Change by at
least 1.5°C.
Check probe and cable connections.
Ensure fluid circuitry is connected
correctly and free from obstruction.
Probe(s) or cable(s) may be
defective. Contact technical support
if problem persists.
1. Ensure pump lids are fully closed
2. Check pumps cable connections
3. Ensure fluid lines/tubing is
properly connected and free from
obstruction/kinks
4. Inspect for fluid drop in burette(s)
and flow from burette through the
probe(s) and back to the burette.
FAULT [E170]
▲
Software Error Detected
Turn AC power OFF-ON to attempt
recovery. Contact technical support
if problem persists.
1. Wait until screen is fully visible
before switching from one mode to
the next
2. Turn unit “OFF” then “ON” to
attempt recovery
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
WARNING [W200]
▲
High Impedance Detected
Probe (X) impedance higher than
recommended maximum (500
Ohms).
1. Check probes and associated
cables
2. Check grounding pad
connections
3. Try new probe(s) and cable(s) if
problem persists
WARNING [W201]
▲
Low Impedance Detected
Possible short circuit between
electrodes, or damaged cannula
insulation.
1. Check probes, cables and
grounding pad connection
2. Try new probe(s) and cable(s) if
problem persists
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3