PAIN MANAGEMENT SYSTEMS
Generator Service Manual
WARNING [W202]
▲
High Impedance Detected
(IDL only)
Check probe and cable connections.
Probe or cable may be defective.
1. Check probe and cable
connections
2. Try new probe(s) and cable(s) if
problem persists
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
WARNING [W203]
▲
Low Impedance Detected
(IDL only)
Check probe and cable connections.
Possible short circuit in probe or
cable.
1. Check probe and cable
connections
2. Try new probe(s) and cable(s) if
problem persists
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
WARNING [W204]
▲
Invalid Temperature Reading
(Standard RF only)
Check probe and cable connections.
Possible defective probe or cable.
Try new probe and cable if problem
persists.
1. Check probe and cable
connections
2. Try new probe(s) and cable(s) if
problem persists
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
WARNING [W205]
▲
Probe B Connected But Disabled
In ADVANCED SETTINGS (TD mode)
Disconnect Probe B or, if desired,
enable 2 PROBES in ADVANCED
SETTINGS.
1. Disconnect probe B
2. Enable 2 probes in ADVANCED
SETTINGS
3. Try new probe(s) and cable(s) if
problem persists
4. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option
WARNING [W206]
▲
Secondary Thermocouple
Connected But Disabled In
ADVANCED SETTINGS
Disconnect Secondary
Thermocouple or, if desired, enable
Secondary Thermocouple in
ADVANCED SETTINGS.
1. Disconnect secondary
thermocouple in RFA ADVANCED
SETTING
2. Enable Secondary Thermocouple
in ADVANCED SETTING
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3