WI-25023
36
PAIN MANAGEMENT SYSTEMS
Generator Service Manual
WARNING [W208]
▲
High Impedance Detected
Check stylet/probe, cable, and
dispersive return electrode
connections. Possible poor tissue
contact.
1. Carefully reposition and stabilize
cannula or introducer
2. Check probes, cable, and
grounding pad connection
3. Try new probe(s) and cable(s) if
problem persists
4. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
WARNING [W209]
▲
Low Impedance Detected
Possible short circuit between
electrodes or damaged introducer
insulation.
WARNING [W210]
▲
High Impedance Detected
Check probe and cable connections.
Possible poor tissue contact.
Probe(s) or cable(s) may be
defective.
1. Carefully reposition and stabilize
cannula or introducer
2. Swap out probe, cable and
grounding pad individually
3. If problem persists, remove probe
from cannula/introducer, wipe off
tip and replace
4. Try new probe(s) and cable(s)
5. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
WARNING [W211]
▲
Select Probe
Multiple Probes Connected.
Multiple probe connectivity not
supported in this mode. Disconnect
extra probes to continue.
1. Disconnect extra probes to
continue
2. Change setting to incorporate
multiple probes
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3