CCD Data Acquisition Module C9266-03/C9266-04
2
2. Setup
2.1 Parts
description
C9266-03/C9266-04 Front Panel
C9266-03 Rear Panel
C9266-04 Rear Panel
(1) CCD
connector
This is a high-density 35-pin MDR connector for the Hamamatsu CCD area image sensor
(S10810, S10811). A bias voltage and clock signal are fed to the CCD through this
connector. Analog video signals from the CCD proportional to X-ray intensity levels and
synchronized timing signals (H-CMOS logic level signals) are input to this connector.
(2) USB
connector
This is an industrial standard USB connector for connecting to a PC. This connector
interface conforms to USB 2.0. Various C9266-03/C9266-04 settings can be made from
the PC through this interface. X-ray image data converted to digital signals are sent to the
PC.
(6)
(2)
(4)
(5)
(3)
(1)
(2)
(3)