Chapter 3. Preparation of System
EQ-77X LDLS® Operation Manual
15
of
31
3.3 Physical Specifications
Dimensions (H x W x D)
Weight
3.4 Utility Requirements
3.5 Environmental Requirements
Operating
Transport
3.6 Remote Interface
Digital Inputs
•
Lamp Head: 128 x 175 x 102 mm (5.0 x 6.9 x 4.0 in)
•
Power Supply Controller: 152 x 250 x 132 mm (6.0 x 9.8 x 5.2 in)
•
Lamp Head: 2.2 kg (4.9 lb)
•
Power Supply Controller: 2.9 kg (6.5 lb)
• Electrical: 100 – 240 V~ +/- 10%, 50/60 Hz, 350 W (Fuse:F5AL250V)
•
◦
Purge Gas: Recommended. Clean dry nitrogen, grade 4.8 or higher, filtered to 5um, 20 psig (0.14 MPa) supply
pressure
Fittings: 4mm push-to-connect
•
◦
Cooling Fluid: Required. ≥ 0.5 liter/minute, 18 – 25 °C, 100 psig (0.69 MPa) max. inlet pressure
Fittings: ¼ inch Swagelok
• Ambient temperature: 15–35 °C.
•
Relative Humidity: non-condensing, 80% max. for temperatures up to 31 °C, decreasing linearly to 50% max. at
35 °C.
•
Pollution Degree 2 (normally only non-conductive pollution; occasional, temporary condensation possible).
•
Overvoltage Category II
•
Indoor use only.
•
Operating altitude 2,000 m max.
•
IP Code IP20
• Temperature: -5–70 °C.
•
Relative Humidity: non-condensing, 95% max.
•
Type: Optocoupler LED
•
Logic: Active High
•
Input voltage: 5VDC
•
Input current: 8 mA