32
16.
Water-cooled type (L, R Series) handling
Please refer to this manual for the basic operation of LCOS-SLM. This section is only an excerpt of how to
use the water-cooled type.
Caution
The heat sink is very heavy. Please use with sufficient care when
moving or moving.
Absolute maximum ratings
Parameter
Value
Unit
WATER pressure
0.3
MPa
When storing the product, make sure to drain the water inside the heat sink and avoid freezing it inside.
Recommended Usage Conditions
Parameter
Recommend condition
Circulating water temperature
25°C
Circulating water quality
Soft water (except for deionized water)
Circulating water flow
0.5 L/min or more
Circulating water hose
Outside diameter φ6mm Inside diameter φ4mm
Name and Description of water-cooled heat sink
Fig.23
water-cooled type
○
WATER connector
It is a connector to insert a hose to flow water to the heat sink. The outside diameter of the applicable
hose is φ6 mm.
○
Stop ring
Stopper for circulating water hose. When you remove the hose, you can remove it by pushing the stop
ring to the body side.
stopping
Water Connector