36
If you want to use the mounting holes on the bottom, you must remove the head with the water-cooled heat
sink (main body) once.
【
Removal - Reconnection Procedure
】
1. Use a screwdriver to loosen and remove the L-shaped fixture fixing screw on the back of the
LCOS-SLM head water-cooled type.
2. Separate the head with water-cooled heat sink (main body) from the L-shaped adapter for
water-cooled type, and attach the L-time adapter to each optical stage.
3. Wipe the front surface of the L-shaped adapter and the back surface of the water-cooled heat sink
with a dry cloth, etc., and attach the L-shaped adapter with the fixing screws.
Caution
Do not screw the L-shaped adapter with foreign matter on the surface.
When securing the screws, tighten the four screws evenly.
Otherwise, the water-cooled heat sink or LCOS-LLM head can be
damaged. These LCOS-SLM head failures are not covered. Please be
careful.
Warning
Do not remove the screws on the back of the head (main body). The
product is not covered by the warranty.
Caution
The heat sink is very heavy. Be careful when removing the L-shaped
adapter. The LCOS-SLM head water-cooled type may be damaged or
damaged if subjected to a strong impact such as dropping. These
LCOS-SLM head failures are not covered. Please be careful.
Fig.26
LCOS-SLM Head Water Cooled Type