15
5.
Go to Step 2: Qualify Laser
Go to Step 2: Qualify Laser and ensure you are
getting readings in all four displays.
6.
Rotate the spindle/L-702 to the INverted
Position.
Rotate the spindle/L-702 to the
INverted
position
(A/C connector panel is facing
toward
you), with
the target in the
NORMal
(cable
down) position. Wait for the
readings to stabilize (approximately
5 seconds). Ensure the bubble level
is in the center of the circle.
Note:
Do NOT rotate the laser in the chuck.
Always rotate the laser and spindle TOGETHER as
one unit.
7.
Click RECORD
.
Click or press
Record
to record the INverted value.
8.
Rotate the spindle/laser to the NORMal
position.
Rotate the spindle/L-702 to the
NORMal
position
(switch panel facing
toward
you with
the level vial
leveled), with the target in the
NORMal
position
(cable down) and wait for the readings to stabilize
(approximately 5 seconds). Ensure the bubble level
is in the
center
of the circle.