  Note:  CURSOR MEASURE can not be called when 

FFT mode is present.  

Depending on the mode (Yt or XY, analog or digital mode) dif-
ferent cursor measure functions can be chosen is this menu, 
regarding the cursor lines and their direction.

The cursor lines and the measurement result are displayed 
after the “Cursors” menu is switched off by pressing the 
MENU OFF pushbutton 

44

. The results of cursor measure-

ments will be displayed by the readout in the top right corner 
of the screen. (e.g. 

Δ

V(CH2):16.6 mV). If a variable control 

was activated, the readout will indicate this by replacing the 
”:“ by a ”

>

“.

Cursor positioning

The cursor lines and symbols can be moved by POSITION 1 

13

 and POSITION 2 

14

 knobs after being activated as cursor 

controls. The POSITION knob function can be selected in the 
“Pos./Scale” menu which can be called by pressing the CH1/2-
CURSOR-MA/REF-ZOOM pushbutton 

15

. In this menu the selec-

tion of ”Cursors“ (long lines) or ”Aux Cursors“ (short lines) or 
other symbols will determine which cursor lines/symbols can 
be moved by the POSITION 1 and 2 controls.    

Menu items

Depending on the operating mode (Analog, Digital, Yt or XY) this 
menu will offer various cursor measuring functions which will 
also affect the cursor lines and their position. The function key 
“Off” switches the cursors off, cause the “CURSORS” menu to 
be left and switches the cursor measuring result display in the 
readout off. 

4.1  Meas.(uring) Type

If this function is activated, the INTENS knob 

2

 can be used 

for measurement selection. In most cases the corresponding 
unit will be automatically displayed. The measuring modes are 
self explaining.   

4.2  Unit 

In the modes ”Ratio X“, ”Ratio Y“ and ”Count“ the INTENS knob 
symbol 

 will be shown in addition to a unit, this may then be 

used to change the unit. 

–    ” rat ” (ratio), display of ratios

 

In this mode the ratios of duty cycles or amplitudes may be 
determined with the CURSORS. The distance between the 
long CURSOR lines is equal to 1.

–   ” % ” (percent), display of percentages

 

The distance between the long CURSOR lines is equal to 100%. 
The result will be determined by the distance of the short 
auxiliary cursor line to the long reference line (lower respec-
tively left), if appropriate with a negative sign.

–   ” ° ” (degree), measurement of degrees

 

The distance between the long CURSOR lines is equal to 360 
degrees and must be exactly as long as a signal period. The 
measurement result will be determined from the distance 
between the reference line to the short auxiliary cursor line. 
If appropriate with a negative sign. For further information 
please consult ”Measurements of phase differences in dual 
channel mode (Yt)“ in the section ”First time operation and 
pre settings“.

   ” 

π

 ”

 

One period of a sine wave is equal to 2 

π

, hence the dis-

tance between the two long CURSOR lines must be set to 
one period. If the distance between the reference line and 

the short CURSOR line equals 1.5 periods, ”3 

π

“ will be 

displayed. If the short cursor line is left of the reference 
line a negative sign will be shown. 

In combination with the measuring mode “Count” the INTENS 
knob symbol 

 is displayed additionally to the unit, indicating 

that it can be determined by the user. 

–   positive Puls

 

The number of such pulses, that are located between the 
vertical cursor lines and crossing the horizontal auxiliary 
cursor line, are displayed. 

– negative 

Puls

 

The number of such pulses, that are located between the 
vertical cursor lines and crossing the horizontal auxiliary 
cursor line, are displayed. 

– rising 

edge

 

The number of rising edges, that are located between the 
vertical cursor lines and crossing the horizontal auxiliary 
cursor line, are displayed. 

–   falling edge

 

The number of falling edges, that are located between the 
vertical cursor lines and crossing the horizontal auxiliary 
cursor line, are displayed. 

4.3 Respect

It may be necessary to determine for which signal channel the 
CURSOR measurement shall be valid. This is indicated by sho-
wing the INTENS knob symbol 

 next to the channel number. 

After selection of the signal source, the CURSOR lines must 
then be positioned to the signal or portions of it displayed by 
this channel.

4.4 Off 

(Cursors Off)

Pressing the function key “Off” switches the “Cursors” menu, 
the cursor lines and the cursor measurement results off.
If only the “Cursors” menu displayed shall be switched off, press 
the MENU OFF pushbutton 

44

.

5

 ANALOG

/

DIGITAL 

(pushbutton)

The color in which the pushbutton lights, indicates the operating 
mode (analog= green, digital = blue). In case of Yt or XY mode no  
settings are affected by switch over. As FFT is only available in 
digital mode, it is switched off after switch over to analog mode 
and the last used Yt mode is present. If COMPONENT TEST 
mode is present (analog mode), switching over to digital mode 
automatically switches over to the last used mode (Yt or XY). 

Y parameters will not be changed by switching. Time base mo-
des will be changed due to the different operation of the time 
bases in both modes. After any switching, time base A will be 
selected. The time base speeds selected will not be affected 
unless they are not available any more, then the maximum 
value will be chosen. 

6

 RUN

/

STOP 

(pushbutton)

 

This pushbutton has several functions: 

6.1  Analog mode, single event capture

The RUN/STOP pushbutton concerns the display of so called 
one time events. To prepare the instrument for this, call the 
“Trigger“ menu by pressing the MODE pushbutton 

20

 and ac-

          C o n t r o l s  a n d  R e a d o u t

Summary of Contents for CombiScope HM2008

Page 1: ...200 MHz Mixed Signal CombiScope HM2008 Manual English...

Page 2: ...irective des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied Normes harmonis es utilis es Sicherheit Safety S curit EN 61010 1 2...

Page 3: ...ajous gures 14 Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode...

Page 4: ...t voltage Time Base 2ns div 50s div Acquisition modes Single Refresh Average Envelope Roll Peak Detect Front USB Stick Connector for Screenshots USB RS 232 optional IEEE 488 or Ethernet USB Signal dis...

Page 5: ...range 0 250 MHz Coupling DC Level control range 10 10 div Horizontal Deflection Analog Time Base Operating modes A ALT alternating A B B Time base A 20 ns div 0 5 s div 1 2 5 sequence Time base B 20 n...

Page 6: ...rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked A A B B C C D D E E T F PUkT PUkT PUk PUk PUk PUk PUk PUk PUkT...

Page 7: ...may be 20 C to 70 C Please note that after exposure to such temperatures or in case of condensation proper time must be allowed until the instrument has reached the permissible temperature resp until...

Page 8: ...V SCALE VAR knob 34 Channel 1 Y de ection coef cient Y variabel and Y scaling setting 17 VOLTS DIV SCALE VAR knob 34 Channel 2 Y de ection coef cient Y variabel and Y scaling setting 18 AUTO MEASURE p...

Page 9: ...selection of PROBE ADJ square wave signal hard ware and software information and details about interface rear side and USB Stick ash drive connector 42 COMPONENT TESTER 2 sockets with 4 mm 47 Connecto...

Page 10: ...st be taken however when charging and discharging a large capacitor DC coupling is preferable with all signals of varying duty cycle otherwise the display will move up and down depending on the duty c...

Page 11: ...ursors may be used for measurement Measurement Adjust the rising portion of the signal to 5 cm Position the rising portion symmetrically to the graticule centre line using both Y and X positioning con...

Page 12: ...e ground connections of probes are especially critical they should be as short as possible and of large size STOP If a probe is to be connected to a BNC connector use a probe tip to BNC adapter If rip...

Page 13: ...mind The oscillosco pe vertical ampli ers are two separate ampli ers and do not constitute a true difference ampli er with as well a high CM rejection as a high permissible CM range Therefore please o...

Page 14: ...and Y signals af ter reading a and b off the screen is possible using the following formulas and a pocket calculator with trigonometric functions This calculation is independent of the signal amplitud...

Page 15: ...ime base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the amp...

Page 16: ...ange the slope Positive or negative slope may be selected in auto or normal trigger modes Also a setting both may be selected which will cause a trigger irrespective of the polarity of the next slope...

Page 17: ...Insert a 100 resistor between the center conductor and the BNC connector If possible the coil should be shielded without creating a shorted winding Alternate trigger This mode is selected with SOURCE...

Page 18: ...time for most of the time the digital will not see the signal Also in analog mode the signal itself will be seen on the screen in real time whereas a digital can only show a reconstruction of the sig...

Page 19: ...er COMPONENT PROBE 40 and COMPONENT TESTER 41 The scope has a built in component tester The test object is connected with 4 mm banana plugs In this mode the Y ampli ers and the time base are turned of...

Page 20: ...pend strongly on the kind of FET With depletion type MOSFETs and all JFETs the channel will conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very...

Page 21: ...ependent single event recording with Yt display Single Readout display sgl 4 Trigger independent continuous recording menu VERT XY with XY display XY Readout display XY 5 XY display of previously in Y...

Page 22: ...value of the sampling interval 1 sampling interval sampling rate 3 The signal frequency value is related to the highest sine wave signal frequency that can be captured with 10 samples per period realt...

Page 23: ...s shows a straight line in the position of the positive peak value and consequently suggest a positive DC voltage An alias may also take the form of a signal of much lower frequency beat frequency bet...

Page 24: ...rogram for USB connection test RS 232 The RS 232 interface has the usual 9 pole SubD connector Via this bidirectional interface the scope can be controlled remo tely or its settings may be transferred...

Page 25: ...ilable submenus In some modes various pushbuttons or INTENS operations are meaningless and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in...

Page 26: ...to leave the menu Thereafter only the signal display is activated and the FOCUS TRACE MENU pushbutton will blink as long as the readout is off Pressing the ashing pushbutton calls the Int Knob menu wh...

Page 27: ...hannel mode Yt in the section First time operation and pre settings One period of a sine wave is equal to 2 hence the dis tance between the two long CURSOR lines must be set to one period If the dista...

Page 28: ...will be lost after turn off Mathematics offers 7 1 Equations set Using the INTENS knob 5 sets of formulas can be selected for editing This way 5 user de ned formula sets may be created Each set of fo...

Page 29: ...he display of the former The display will remain on screen until the next acquisition is started by the trigger and causes a new recording This mode is available over the full time base range 50 s cm...

Page 30: ...size is 0 25 s between the samples at a time base setting of 50 s cm The advantage of Peak Detect is that signals are sampled with the highest possible sampling rate to reduce the gap size so that eve...

Page 31: ...RE1 RE1 the signal will be displayed twice on the same spot Associated settings Pressing this function key causes the oscilloscope to take over the settings parameter stored when the signal was store...

Page 32: ...s will remain The digital modes Roll Envelope or Average will be changed to refresh mode AUTOSET will further set the intensity to an average value if it was set too low If a menu was opened it will b...

Page 33: ...control if Yt mode time base mode is present and the CH1 2 CURSOR MA REF ZOOM pushbutton 15 is not lit 14 1 2 Y Position of the FFT display only in digital mode measured with channel 2 Position 2 allo...

Page 34: ...he CH1 menu the de ection coef cient becomes calibrated the CH1 VAR pushbutton 31 does not light any longer and the VOLTS DIV SCALE VAR knob 16 returns to 1 2 5 sequence 16 3 Scaling the FFT display d...

Page 35: ...at V rms the range is from 5mV cm to 20V cm also switched in 1 2 5 sequence Please note that in contrast to Yt and XY mode signal display no peak to peak values but rms values are displayed 18 AUTO ME...

Page 36: ...me base set B trigger to positive or negative slope Thereafter the LEVEL A B control will control the time base B trigger a second trigger point symbol will be shown and marked with B 19 2 FFT mode Th...

Page 37: ...mpli er bandwidth and consequently less trigger signal noise Readout Tr Source Slope AC or DC NR 21 1 2 Slope SLOPE determines whether the rising or falling portion of a signal shall trigger the level...

Page 38: ...5 External In this mode the trigger signal comes from the AUXILIARY INPUT TRIG EXT 38 Readout Tr ext Slope Filter 22 1 6 AC Line The trigger signal is taken from the line mains supply which feeds the...

Page 39: ...ither reference nor mathematic signals are displayed The expanded portion of the signal will be shown on the normal display as an intensi ed sector The length of this sector is dependent upon the sett...

Page 40: ...ave form become digitised and at least show spectrum displays spectra that do not exist in reality If the sampling rate is too low the readout displays ALS if the signal is too high overrange will be...

Page 41: ...nly B The TIME DIV SCA LE VAR knob then only affects time base B By aid of the MAG x10 pushbutton 29 the signal display can be expanded 10 fold reducing the time de ection coef cient by factor 10 30 1...

Page 42: ...sensitivity setting is for AC and is for DC 31 1 1 DC coupling The signal will be directly coupled from the BNC connector via the attenuator to the vertical ampli er The input resistance is 1 M in all...

Page 43: ...DUAL trace alt chop In dual trace mode both channels are turned on and the para meters of both are shown in the readout Between the sensi tivity indications there is an indication whether alternate al...

Page 44: ...It is recommended rst to look at the signals in DUAL mode and to set the sampling rate such that at least one signal period will be displayed Then XY should be selected In XY digital mode both channe...

Page 45: ...he VOLTS DIV SCALE VAR knob 17 returns to 1 2 5 sequence switching and CH2 to calibrated condition 34 INPUT CH1 BNC connector This is the CH1 signal input connector In Yt mode it is a Y input in XY mo...

Page 46: ...accuracy of the FFT calculations in the following averageing process Numbers between 2 and 512 can be chosen by aid of the INTENS 2 knob which are displayed by the readout e g avg 256 The accuracy in...

Page 47: ...ive is connected with the USB Stick connector 43 41 4 1 USB Stick Save Front Panel current oscilloscope parameters The current parameters are saved on the USB ash drive by pressing the function key Sa...

Page 48: ...D 63533 Mainhausen A Rohde Schwarz Company Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg com w w w h a m e g c o m 41 2008 00E1 authori...

Reviews: