background image

29

Subject to change without notice

After MA the next step causes  an arrow symbol to be displayed 
in the “Mathematics Edit“ menu under item “Edit“. It is a hint 
about another submenu described under item 7.2.5.

7.2.5

 Operand selection by Constant Editor 

In the CW position of the INTENS knob 

2

 an additional item 

”Edit“ and an arrow symbol is displayed. Pressing the associated 
function button opens the sub submenu ”Edit Edit“. This enables 
you to choose a number, its ”Dec.Point“ (decimal point) and its 
”Prefix“ of the ”Unit“, all with the INTENS knob 

2

 . The unit need 

not be selected: it is only shown as a reminder. A constant selec-
ted as an operand will be displayed as a unit in the menu.

7.3 Display

STOP

 Attention! 

 

“Mathematics signals“ (MA1 to MA5), “Reference 
signals“ (RE1 to RE9) and “Logic signals“ (LC0 to 
LC3) cannot be displayed at the same time. Display-
ing “Mathematics signals“ automatically switches 
the display of “Reference signals“ and “Logic 
signals“ off, and vice versa. 

 
The function ”Display“ is available twice in the menu and may 
be switched on or off in any combination. This allows you to 
display: no result, one result, or the result of two equations 
as signals. The displays will come forward upon leaving the 
MATHEMATICS menu. Also the designation of the equation (e.g. 
MA2) will be shown. Select the equations to be displayed with 
the INTENS knob 

2

The mathematics signal is automatically scaled, this is indepen-
dent of the graticule, of Y and time base parameters, hence the 
scale will not be shown. The measurement of the signal ampli-
tudes must be performed using the CURSOR (V to GND) after 
the REFERENCE (e.g. MA2) of the CURSOR to the ”mathematics 
signal“ and its scale is established (CURSOR MEASURE 

4

pushbutton 

>

Cursors 

>

reference 

>

 e.g. MA2). The readout may 

then display e.g.: ”V(MA2): 900 mV“. 

In combination division and the constant 0, no result will be 
displayed. The mathematics function will only be calculated 
and displayed if possible. As the calculation is made in real time 
conditions, a new calculation requires that the channels are 
activated and new valid data are present (e.g. trigger conditions 
are met). An error message will not be output.
 

7.4 Units 

Each function ”Display“ will be associated with a function 
”Unit“ which can be selected with INTENS knob 

2

 and will be 

attached to the result. This enables you to allocate a unit to 
each mathematics curve. In the case of cursor and automatic 
measurement related to mathematic curves it will be indicated 
by the readout as the source.  

8

 

Acquire 

(pushbutton)

 

(digital mode only, in FFT mode without effect)

This pushbutton opens the menu ACQUIRE which offers these 
modes: 

8.1  Normal (Refresh) – Capture/Display

In this mode repetitive signals can be recorded and displayed, 
as in analog mode; the display shows rfr. The current signal 
capture can be stopped (pushbutton lit) or started (pushbutton 
not lit) by the RUN/STOP pushbutton 

6

.   

A trigger will start a new acquisition which will overwrite the 
display of the former. The display will remain on screen until 

the next acquisition is started by the trigger and causes a new 
recording. This mode is available over the full time base range 
(50 s/cm to 2 ns/cm).

STOP

 Please 

note: 

 

At 20 ms and smaller time defl ection coeffi cients 
the signal display always starts at the screen left. 
After switching to a time defl ection coeffi cient of 
50 ms or higher, the capture starts at once but 
under these conditions the signal display starts 
at the trigger point, which is placed without delay 
(Readout: “Tt:0s“) at the screen horizontal center. 
The second capture starts at the screen left. 

This behaviour is relatively meaningless for small time defl ec-
tion coeffi cients. In the case of high time defl ection coeffi cients 
combined with high Post Trigger times, the instrument seems 
to show no reaction. The resulting uncertainty can be avoided 
by activating the STATUS display (see item 10.4.4 Status Acq. 
(acquisition) 

The following example describes the long waiting times caused 
by the 1 M Byte RAM: 

With the time base setting at 50 s/cm and the trigger point  set 
to the utmost left position by the HORIZONTAL control 

27

, the 

readout will indicate ”Tt:1.85ks“. This means that 1,600 seconds 
must elapse until the trace will become visible at the screen left 
and after another 250 s it will have reached the screen centre 
(1,600 s + 250 s = 1.85 ks). 

After the capture has been completed, the new signal curve 
overwrites that previously recorded, after a trigger event started 
the capture and the waiting time elapsed.

8.2 Envelope capture/display 

Envelope is a special refresh mode, the readout will show 
”env“. Also in this mode there must be sufficient signal for 
triggering. 

In contrast to the refresh mode the results of several captures 
will be examined and the maxima and minima stored, the enve-
lope of the signal will then be displayed if it changes in amplitude 
or/and frequency. Also any jitter will be shown. 

Also in this mode pressing RUN/STOP 

6

 will stop the 

acquisition, indicated by STOP constantly lit. After pressing the 
pushbutton again, the formerly stored signals will be erased 
and the envelope calculation starts anew. In order to prevent an 
accidental turning on of this mode, operating any control which 
influences the signal display will automatically switch envelope 
off. Because this mode requires many signal repetitions and 
acquisitions it is not compatible with single sweep/acquisition. 
AUTO or normal trigger modes must be selected. 

8.3 Average mode capture/display 

This is also a special mode within the refresh mode. Also here 
signal repetitions are needed. 

The weighting of each acquisition can be selected with ”Ave-
rage“ in the menu, any number between 2 and 512 may be 
chosen using the INTENS knob 

2

. The readout will show e.g. 

”avg#512“. 

The higher the number of acquisitions averaged, the lower the 
contribution of a single acquisition will be and the longer the 
averaging will take. Averaging is a means to increase the ac-
curacy in spite of the 8 bit converters, it is an exchange of time 
against accuracy. Noise will be reduced by averaging. 

          C o n t r o l s  a n d  R e a d o u t

Summary of Contents for CombiScope HM2008

Page 1: ...200 MHz Mixed Signal CombiScope HM2008 Manual English...

Page 2: ...irective des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied Normes harmonis es utilis es Sicherheit Safety S curit EN 61010 1 2...

Page 3: ...ajous gures 14 Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode...

Page 4: ...t voltage Time Base 2ns div 50s div Acquisition modes Single Refresh Average Envelope Roll Peak Detect Front USB Stick Connector for Screenshots USB RS 232 optional IEEE 488 or Ethernet USB Signal dis...

Page 5: ...range 0 250 MHz Coupling DC Level control range 10 10 div Horizontal Deflection Analog Time Base Operating modes A ALT alternating A B B Time base A 20 ns div 0 5 s div 1 2 5 sequence Time base B 20 n...

Page 6: ...rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked A A B B C C D D E E T F PUkT PUkT PUk PUk PUk PUk PUk PUk PUkT...

Page 7: ...may be 20 C to 70 C Please note that after exposure to such temperatures or in case of condensation proper time must be allowed until the instrument has reached the permissible temperature resp until...

Page 8: ...V SCALE VAR knob 34 Channel 1 Y de ection coef cient Y variabel and Y scaling setting 17 VOLTS DIV SCALE VAR knob 34 Channel 2 Y de ection coef cient Y variabel and Y scaling setting 18 AUTO MEASURE p...

Page 9: ...selection of PROBE ADJ square wave signal hard ware and software information and details about interface rear side and USB Stick ash drive connector 42 COMPONENT TESTER 2 sockets with 4 mm 47 Connecto...

Page 10: ...st be taken however when charging and discharging a large capacitor DC coupling is preferable with all signals of varying duty cycle otherwise the display will move up and down depending on the duty c...

Page 11: ...ursors may be used for measurement Measurement Adjust the rising portion of the signal to 5 cm Position the rising portion symmetrically to the graticule centre line using both Y and X positioning con...

Page 12: ...e ground connections of probes are especially critical they should be as short as possible and of large size STOP If a probe is to be connected to a BNC connector use a probe tip to BNC adapter If rip...

Page 13: ...mind The oscillosco pe vertical ampli ers are two separate ampli ers and do not constitute a true difference ampli er with as well a high CM rejection as a high permissible CM range Therefore please o...

Page 14: ...and Y signals af ter reading a and b off the screen is possible using the following formulas and a pocket calculator with trigonometric functions This calculation is independent of the signal amplitud...

Page 15: ...ime base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the amp...

Page 16: ...ange the slope Positive or negative slope may be selected in auto or normal trigger modes Also a setting both may be selected which will cause a trigger irrespective of the polarity of the next slope...

Page 17: ...Insert a 100 resistor between the center conductor and the BNC connector If possible the coil should be shielded without creating a shorted winding Alternate trigger This mode is selected with SOURCE...

Page 18: ...time for most of the time the digital will not see the signal Also in analog mode the signal itself will be seen on the screen in real time whereas a digital can only show a reconstruction of the sig...

Page 19: ...er COMPONENT PROBE 40 and COMPONENT TESTER 41 The scope has a built in component tester The test object is connected with 4 mm banana plugs In this mode the Y ampli ers and the time base are turned of...

Page 20: ...pend strongly on the kind of FET With depletion type MOSFETs and all JFETs the channel will conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very...

Page 21: ...ependent single event recording with Yt display Single Readout display sgl 4 Trigger independent continuous recording menu VERT XY with XY display XY Readout display XY 5 XY display of previously in Y...

Page 22: ...value of the sampling interval 1 sampling interval sampling rate 3 The signal frequency value is related to the highest sine wave signal frequency that can be captured with 10 samples per period realt...

Page 23: ...s shows a straight line in the position of the positive peak value and consequently suggest a positive DC voltage An alias may also take the form of a signal of much lower frequency beat frequency bet...

Page 24: ...rogram for USB connection test RS 232 The RS 232 interface has the usual 9 pole SubD connector Via this bidirectional interface the scope can be controlled remo tely or its settings may be transferred...

Page 25: ...ilable submenus In some modes various pushbuttons or INTENS operations are meaningless and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in...

Page 26: ...to leave the menu Thereafter only the signal display is activated and the FOCUS TRACE MENU pushbutton will blink as long as the readout is off Pressing the ashing pushbutton calls the Int Knob menu wh...

Page 27: ...hannel mode Yt in the section First time operation and pre settings One period of a sine wave is equal to 2 hence the dis tance between the two long CURSOR lines must be set to one period If the dista...

Page 28: ...will be lost after turn off Mathematics offers 7 1 Equations set Using the INTENS knob 5 sets of formulas can be selected for editing This way 5 user de ned formula sets may be created Each set of fo...

Page 29: ...he display of the former The display will remain on screen until the next acquisition is started by the trigger and causes a new recording This mode is available over the full time base range 50 s cm...

Page 30: ...size is 0 25 s between the samples at a time base setting of 50 s cm The advantage of Peak Detect is that signals are sampled with the highest possible sampling rate to reduce the gap size so that eve...

Page 31: ...RE1 RE1 the signal will be displayed twice on the same spot Associated settings Pressing this function key causes the oscilloscope to take over the settings parameter stored when the signal was store...

Page 32: ...s will remain The digital modes Roll Envelope or Average will be changed to refresh mode AUTOSET will further set the intensity to an average value if it was set too low If a menu was opened it will b...

Page 33: ...control if Yt mode time base mode is present and the CH1 2 CURSOR MA REF ZOOM pushbutton 15 is not lit 14 1 2 Y Position of the FFT display only in digital mode measured with channel 2 Position 2 allo...

Page 34: ...he CH1 menu the de ection coef cient becomes calibrated the CH1 VAR pushbutton 31 does not light any longer and the VOLTS DIV SCALE VAR knob 16 returns to 1 2 5 sequence 16 3 Scaling the FFT display d...

Page 35: ...at V rms the range is from 5mV cm to 20V cm also switched in 1 2 5 sequence Please note that in contrast to Yt and XY mode signal display no peak to peak values but rms values are displayed 18 AUTO ME...

Page 36: ...me base set B trigger to positive or negative slope Thereafter the LEVEL A B control will control the time base B trigger a second trigger point symbol will be shown and marked with B 19 2 FFT mode Th...

Page 37: ...mpli er bandwidth and consequently less trigger signal noise Readout Tr Source Slope AC or DC NR 21 1 2 Slope SLOPE determines whether the rising or falling portion of a signal shall trigger the level...

Page 38: ...5 External In this mode the trigger signal comes from the AUXILIARY INPUT TRIG EXT 38 Readout Tr ext Slope Filter 22 1 6 AC Line The trigger signal is taken from the line mains supply which feeds the...

Page 39: ...ither reference nor mathematic signals are displayed The expanded portion of the signal will be shown on the normal display as an intensi ed sector The length of this sector is dependent upon the sett...

Page 40: ...ave form become digitised and at least show spectrum displays spectra that do not exist in reality If the sampling rate is too low the readout displays ALS if the signal is too high overrange will be...

Page 41: ...nly B The TIME DIV SCA LE VAR knob then only affects time base B By aid of the MAG x10 pushbutton 29 the signal display can be expanded 10 fold reducing the time de ection coef cient by factor 10 30 1...

Page 42: ...sensitivity setting is for AC and is for DC 31 1 1 DC coupling The signal will be directly coupled from the BNC connector via the attenuator to the vertical ampli er The input resistance is 1 M in all...

Page 43: ...DUAL trace alt chop In dual trace mode both channels are turned on and the para meters of both are shown in the readout Between the sensi tivity indications there is an indication whether alternate al...

Page 44: ...It is recommended rst to look at the signals in DUAL mode and to set the sampling rate such that at least one signal period will be displayed Then XY should be selected In XY digital mode both channe...

Page 45: ...he VOLTS DIV SCALE VAR knob 17 returns to 1 2 5 sequence switching and CH2 to calibrated condition 34 INPUT CH1 BNC connector This is the CH1 signal input connector In Yt mode it is a Y input in XY mo...

Page 46: ...accuracy of the FFT calculations in the following averageing process Numbers between 2 and 512 can be chosen by aid of the INTENS 2 knob which are displayed by the readout e g avg 256 The accuracy in...

Page 47: ...ive is connected with the USB Stick connector 43 41 4 1 USB Stick Save Front Panel current oscilloscope parameters The current parameters are saved on the USB ash drive by pressing the function key Sa...

Page 48: ...D 63533 Mainhausen A Rohde Schwarz Company Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg com w w w h a m e g c o m 41 2008 00E1 authori...

Reviews: