background image

14

Subject to change without notice

After completion of the HF-adjustment, the signal amplitude
displayed on the CRT screen should have the same value
as during the 1kHz adjustment.

Probes other than those mentioned above, normally
have a larger tip diameter and may not fit into the
calibrator outputs. Whilst it is not difficult for an
experienced operator to build a suitable adapter, it
should be pointed out that most of these probes have
a slower risetime with the effect that the total bandwidth
of scope together with probe may fall far below that of
the HM303. Furthermore, the HF-adjustment feature is
nearly always missing so that waveform distortion can
not be entirely excluded.

The adjustment sequence must be followed in the order
described, i.e. first at 1kHz, then at 1MHz. The calibrator
frequencies should not be used for timebase calibration.
The pulse duty cycle deviates from 1:1 ratio.

Prerequisites for precise and easy probe adjustments, as
well as checks of deflection coefficients, are straight
horizontal pulse tops, calibrated pulse amplitude, and
zero-potential at the pulse base. Frequency and duty cycle
are relatively uncritical. For interpretation of transient
response, fast pulse risetimes and low-impedance
generator outputs are of particular importance.

Providing these essential features, as well as switch-
selectable output-frequencies, the calibrator of the HM303
can, under certain conditions, replace expensive
squarewave generators when testing or compensating
wideband-attenuators or -amplifiers. In such a case, the
input of an appropriate circuit will be connected to one of
the CAL.-outputs via a suitable probe.

The voltage provided at a high-impedance input (1M

II15-

50pF) will correspond to the division ratio of the probe
used (10:1 = 20mV

pp

, 100:1 = also 20mV

pp

 from 2V

output). Suitable probes are HZ51, 52, 53, and 54.

Operating modes of the vertical amplifiers

The vertical amplifier is set to the desired operating mode
by using the 3 pushbuttons (CH I/II, DUAL and ADD) in the
Y field of the front panel. For 

Mono

 mode all 3 buttons

must be in their released positions; only channel I can then
be operated. The button CH I/II-TRIG.I/II must be depressed
in mono mode for Channel II. The internal triggering is
simultaneously switched over to Channel II with this
button.

If the 

DUAL

 button is depressed, both channels are

working. Two signals can be displayed together in this
button position (alternate mode) if the time-base setting
and the repetition frequency of the signal are suited. This
mode is not suitable for displaying very slow-running
processes. The display then flickers too much or it appears
to jump. If the 

ADD

 button is depressed 

in addition

 to

DUAL

, both channels are switched over constantly at a

high frequency within a sweep period (

CHOP

 mode). Low

frequency signals 

below 1kHz, or with periods longer

than 1ms

 are then also displayed without flicker. CHOP

mode is not recommended for signals with higher repetition
frequencies.

If only the 

ADD

 button is depressed, the signals of both

channels are algebraically added (±I ±II). Whether the
resulting display shows the 

sum

 or 

difference

 is dependent

on the phase relationship or the polarity of the signals 

and

on the positions of the 

INVERT

 buttons.

In-phase input voltages:

Both 

INVERT CH.I 

and INVERT CH.II

 buttons

released or depressed = sum.

Only one INVERT button depressed = difference.

Antiphase input voltages:

Both 

INVERT

 buttons released or depressed

= difference.

INVERT CH.I 

or INVERT CH.II

 button depressed = sum.

In the 

ADD

 mode the vertical display position is dependent

upon the 

Y-POS.

 setting of 

both

 channels. The same

attenuator switch position is normally used for both
channels with algebraic addition.

Please note that the Y-POS. settings are added too but are
not affected by the INVERT pushbuttons.

Differential measurement

 techniques allow direct

measurement of the voltage drop across floating
components (both ends above ground). Two identical
probes should be used for both vertical inputs. In order to
avoid ground loops, use a separate ground connection and
do not use the probe ground leads or cable shields.

incorrect 

incorrect

correct

Adjustment
1MHz

Summary of Contents for HM 303-4

Page 1: ...ge 8 Amplitude Measurements 8 Time Measurements 9 Connection of Test Signal 10 First Time Operation 12 Trace Rotation TR 12 Probe compensation and use 12 Operating Modes of the Y Amplifier 14 X Y Oper...

Page 2: ...ual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection to a cable Basica...

Page 3: ...WG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied...

Page 4: ...nal Features Component Tester 1kHz 1MHz Calibrator OSCILLOSCOPES Specifications Vertical Deflection Operating modes Channel I or II separate both Channels alternated or chopped Chopper frequency appro...

Page 5: ...prox 500M DC 1MHz max 15kV DC peakAC HZ47 Viewing Hood for Oscilloscopes HM205 408 604 1 2 1005 and 1007 HZ48 Viewing Hood for Oscilloscopes 303 304 305 604 3 and 1004 during transpor tation of an osc...

Page 6: ...is standard It has left the factory in a safe condition This instruction manual contains important information and warnings which have to be followed by the user to ensure safe operation and to retain...

Page 7: ...tment designation isstatedforpossiblequeries thishelpstowards speeding up the processing of guarantee claims Maintenance Variousimportantpropertiesoftheoscilloscopeshouldbe carefully checked at certai...

Page 8: ...However forsignalmagnitudesandvoltage designations in oscilloscope measurements the peak to peak voltage Vpp value is applied The latter corresponds to the real potential difference between the most...

Page 9: ...he attenuator series resistor will break down causingdamagetotheinputoftheoscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal...

Page 10: ...must be divided by 10 The fall time of a pulse can also be measured by using this method The following figure shows correct positioning of the oscilloscope trace for accurate risetime measurement tr...

Page 11: ...obecompensation page M7 Standard attenuator probes on the oscilloscope normally reduce its bandwidth and increase the rise time In all cases where the oscilloscope bandwidth must be fully utilized e g...

Page 12: ...ion required The HM303 accepts all signals from DC direct voltage up to a frequency of at least 30MHz 3dB For sinewave voltages the upper frequency limit will be 50MHz 6dB However in this higher frequ...

Page 13: ...should then be 4 div 0 12div 3 During this adjustment the signal edges will remain invisible Adjustment at 1MHz Probes HZ51 52 and 54 can also be HF compensated They incorporate resonance de emphasing...

Page 14: ...the 3 pushbuttons CH I II DUAL and ADD in the Y field of the front panel For Mono mode all 3 buttons mustbeintheirreleasedpositions onlychannelIcanthen beoperated ThebuttonCHI II TRIG I IImustbedepres...

Page 15: ...in the X Y mode can exceed an angle of 3 above 120 kHz It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the...

Page 16: ...o buttons Y CH I 20mV div AC TIME DIV 0 2ms div Triggering NORMAL with LEVEL setting internal or external triggering Figure 2 Amplitude modulated oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms If the...

Page 17: ...ing mode it is possible to trigger at any amplitude point of a signal edge even with very complex signal shapes by adjusting the LEVEL control Its adjusting range is directly dependent on the display...

Page 18: ...eringcanalsobeusefultodisplaysignalsbelowthe triggerthreshold lessthan0 5div Itisthereforeparticularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in...

Page 19: ...iodes marked OVERSCAN which are located between the attenuators Should one LED illuminate without an input signal this means that the respective vertical positioning control has been improperly adjust...

Page 20: ...orizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approxi mately evaluated The determination of actual values will come with...

Page 21: ...n by 180 degrees round about the center point of the scope graticule In Circuit Tests Caution During in circuit tests make sure the circuit is dead No power from mains line or battery and no signal in...

Page 22: ...ort circuit Resistor 510 Junction B C Junction B E Mains transformer prim Capacitor33 F Junction E C FET Z diode below 7V Z diode beyond 7V Diode paralleled by 680 2 Diodes antiparallel Silicon diode...

Page 23: ...n in the edge zone of the screen must be accepted It is limited by standards of the CRT manufacturer The same is valid for tolerances of the orthogonality the undeflected spot position the non lineari...

Page 24: ...anneldisplaywiththe CHI IIbutton is unnecessary it is contained indirectly in the tests above stated input of the vertical amplifier e g using a x1 probe the displayed signal in the50mV div position v...

Page 25: ...splay should not shift horizontally during a change of the trigger coupling from AC to DC with a sine wave signal without DC offset If both vertical inputs are AC coupled to the same signal and both t...

Page 26: ...LDOFF knob Component Tester After pressing the COMP TESTER button a horizontal straight line has to appear immediately when theCOMP TESTER socket is open The length of this trace should be approx 8div...

Page 27: ...d from all voltage sources Normally the capacitors are discharged approx 6 seconds after switching off However with a defective instrument an interruption oftheloadisnotimpossible Therefore afterswitc...

Page 28: ...g Checktoseethatallcircuitboardconnections are making good contact and are not shorting to an adjacent circuit Especially inspect the connections between the PCBs to front chassis parts to CRT PCB to...

Page 29: ...ring TRIG EXT button depressed sync signal 0 3Vpp to 3Vpp to TRIG INP socket Line triggering TRIG selector switch to Select trigger coupling with TRIG selector switch Trigger frequency ranges AC 20Hz...

Page 30: ...nector horizontal deflection in X Y mode Y POS II Controls vertical position knob of channel II display Inoperative in X Y mode POWER Turns scope on and off pushbutton LED LED indicates operating cond...

Page 31: ...1 3 5 6 7 8 9 14 15 16 17 2 4 10 11 12 13 20 21 22 23 24 25 26 27 28 30 31 34 36 37 38 33 32 18 29 19 35...

Reviews: