background image

Subject to change without notice

27

General

The following instructions are intended as an aid for the
electronic technician, who is carrying out readjustments
on the HM303, if the nominal values do not meet the
specifications. These instructions primarily refer to those
faults, which were found after using the Test Instructions.
However, this work should only be carried out by properly
qualified personnel. For any further technical information
call or write to HAMEG. Addresses are provided at the
back of the manual. It is recommended to use only the
original packing material, should the instrument be shipped
to  for service or repair (see also Warranty, page M2).

Instrument Case Removal

The rear cover can be taken off after unplugging the power
cords triple-contact connector and after two cross recessed
pan head screws (M4x30mm) with two washers on it
have been removed. While the instrument case is firmly
held, the entire chassis with its front panel can withdrawn
forward. When the chassis is inserted into the case later
on, it should be noticed that the case has to fit under the
flange of the front panel. The same applies for the rear of
the case, on which the rear cover is put.

Caution

During opening or closing of the case, the instrument
must be disconnected from all power sources for
maintenance work or a change of parts or
components. If a measurement, trouble-shooting, or
an adjustment is unavoidable, this work must be
done by a specialist, who is familiar with the risk
involved.

When the instrument is set into operation after the
case has been removed, attention must be paid to the
acceleration voltage for the CRT  –2025V  and to the
operating voltages for both final amplifier stages
185V and 141V. Potentials of these voltages are on
the PS-Board, the CRT-PCB, on the upper and lower
PCBs. Such potentials are moreover on the checkpoint
strips on the upper and lower horizontal PCBs. They
are highly dangerous and therefore precautions must
be taken. It should be noted furthermore that shorts
occuring on different points of the CRT high voltage
and unblanking circuitry will definitely damage some
semiconductors and the opto-coupler. For the same
reason it is very risky to connect capacitors to these
points while the instrument is on.

Capacitors in the instrument may still be charged,
even when the instrument is disconnected from all
voltage sources. Normally, the capacitors are
discharged approx. 6 seconds after switching off.
However, with a defective instrument an interruption

of the load is not impossible. Therefore, after switching
off, it is recommended to connect one by one all
terminals of the check strips on the upper PCB across
1k

 

to ground (chassis) for a period of 1 second.

Handling of the CRT needs utmost caution. The glass
bulb must not be allowed  under any circumstances
to come into contact with hardened tools, nor should
it undergo local superheating (e.g. by soldering iron)
or local undercooling (e.g. by cryogenic-spray). We
recommend the wearing of safety goggles (implosion
danger).

The complete instrument (with case closed and PO-
WER button depressed) is  after each intervention
undergo a voltage test with 2200V, DC, between
accessible parts to both mains/line supply terminals.
This test is dangerous and requires an adequately
trained specialist

.

Operating Voltages

All operating voltages (+6.3V, +12V, –12V, +141V, +185V,
–2025V) are stabilized by the switch mode power supply.
The +12V supply is further stabilized and used as a
reference voltage for –12V stabilisation. These different
operating voltages are fixed voltages, except the +12V,
which can be adjusted. The variation of the fixed voltages
greater than 5% from the nominal value indicates a fault.
Measurements of the high voltage may only be
accomplished by the use of a sufficient highly resistive
voltmeter (>10M

). You must make absolutely sure that

the electric strength of the voltmeter is sufficiently high.
It is recommended to check the ripple and also the
interaction from other possible sources. Excessive values
might be very often the reason for incomprehensible
faults.

Maximum and Minimum Brightness

Two variable resistors (220k

 and 470k

), located on the

switch mode power supply PCB, are used for these
adjustment procedures (see Adjusting Plan). They may only
be touched by a properly insulating screwdriver (Caution!
High voltage!). The adjustments may possibly have to be
repeated, because the functions of both variable resistors
are dependent on each other. Correct adjustment is achieved,
when the trace can be blanked while 

XY

 pushbutton is

depressed and, in addition, when the requirement described
in the Test Instructions are met.

Astigmatism control

The ratio of vertical and horizontal sharpness can be
adjusted by the variable resistor of 47k

, located on the

lower PCB (see Adjusting Plan). As a precaution however,
the voltage for the vertical deflecting plates (approx.

Service Instructions

Summary of Contents for HM 303-4

Page 1: ...ge 8 Amplitude Measurements 8 Time Measurements 9 Connection of Test Signal 10 First Time Operation 12 Trace Rotation TR 12 Probe compensation and use 12 Operating Modes of the Y Amplifier 14 X Y Oper...

Page 2: ...ual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection to a cable Basica...

Page 3: ...WG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied...

Page 4: ...nal Features Component Tester 1kHz 1MHz Calibrator OSCILLOSCOPES Specifications Vertical Deflection Operating modes Channel I or II separate both Channels alternated or chopped Chopper frequency appro...

Page 5: ...prox 500M DC 1MHz max 15kV DC peakAC HZ47 Viewing Hood for Oscilloscopes HM205 408 604 1 2 1005 and 1007 HZ48 Viewing Hood for Oscilloscopes 303 304 305 604 3 and 1004 during transpor tation of an osc...

Page 6: ...is standard It has left the factory in a safe condition This instruction manual contains important information and warnings which have to be followed by the user to ensure safe operation and to retain...

Page 7: ...tment designation isstatedforpossiblequeries thishelpstowards speeding up the processing of guarantee claims Maintenance Variousimportantpropertiesoftheoscilloscopeshouldbe carefully checked at certai...

Page 8: ...However forsignalmagnitudesandvoltage designations in oscilloscope measurements the peak to peak voltage Vpp value is applied The latter corresponds to the real potential difference between the most...

Page 9: ...he attenuator series resistor will break down causingdamagetotheinputoftheoscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal...

Page 10: ...must be divided by 10 The fall time of a pulse can also be measured by using this method The following figure shows correct positioning of the oscilloscope trace for accurate risetime measurement tr...

Page 11: ...obecompensation page M7 Standard attenuator probes on the oscilloscope normally reduce its bandwidth and increase the rise time In all cases where the oscilloscope bandwidth must be fully utilized e g...

Page 12: ...ion required The HM303 accepts all signals from DC direct voltage up to a frequency of at least 30MHz 3dB For sinewave voltages the upper frequency limit will be 50MHz 6dB However in this higher frequ...

Page 13: ...should then be 4 div 0 12div 3 During this adjustment the signal edges will remain invisible Adjustment at 1MHz Probes HZ51 52 and 54 can also be HF compensated They incorporate resonance de emphasing...

Page 14: ...the 3 pushbuttons CH I II DUAL and ADD in the Y field of the front panel For Mono mode all 3 buttons mustbeintheirreleasedpositions onlychannelIcanthen beoperated ThebuttonCHI II TRIG I IImustbedepres...

Page 15: ...in the X Y mode can exceed an angle of 3 above 120 kHz It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the...

Page 16: ...o buttons Y CH I 20mV div AC TIME DIV 0 2ms div Triggering NORMAL with LEVEL setting internal or external triggering Figure 2 Amplitude modulated oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms If the...

Page 17: ...ing mode it is possible to trigger at any amplitude point of a signal edge even with very complex signal shapes by adjusting the LEVEL control Its adjusting range is directly dependent on the display...

Page 18: ...eringcanalsobeusefultodisplaysignalsbelowthe triggerthreshold lessthan0 5div Itisthereforeparticularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in...

Page 19: ...iodes marked OVERSCAN which are located between the attenuators Should one LED illuminate without an input signal this means that the respective vertical positioning control has been improperly adjust...

Page 20: ...orizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approxi mately evaluated The determination of actual values will come with...

Page 21: ...n by 180 degrees round about the center point of the scope graticule In Circuit Tests Caution During in circuit tests make sure the circuit is dead No power from mains line or battery and no signal in...

Page 22: ...ort circuit Resistor 510 Junction B C Junction B E Mains transformer prim Capacitor33 F Junction E C FET Z diode below 7V Z diode beyond 7V Diode paralleled by 680 2 Diodes antiparallel Silicon diode...

Page 23: ...n in the edge zone of the screen must be accepted It is limited by standards of the CRT manufacturer The same is valid for tolerances of the orthogonality the undeflected spot position the non lineari...

Page 24: ...anneldisplaywiththe CHI IIbutton is unnecessary it is contained indirectly in the tests above stated input of the vertical amplifier e g using a x1 probe the displayed signal in the50mV div position v...

Page 25: ...splay should not shift horizontally during a change of the trigger coupling from AC to DC with a sine wave signal without DC offset If both vertical inputs are AC coupled to the same signal and both t...

Page 26: ...LDOFF knob Component Tester After pressing the COMP TESTER button a horizontal straight line has to appear immediately when theCOMP TESTER socket is open The length of this trace should be approx 8div...

Page 27: ...d from all voltage sources Normally the capacitors are discharged approx 6 seconds after switching off However with a defective instrument an interruption oftheloadisnotimpossible Therefore afterswitc...

Page 28: ...g Checktoseethatallcircuitboardconnections are making good contact and are not shorting to an adjacent circuit Especially inspect the connections between the PCBs to front chassis parts to CRT PCB to...

Page 29: ...ring TRIG EXT button depressed sync signal 0 3Vpp to 3Vpp to TRIG INP socket Line triggering TRIG selector switch to Select trigger coupling with TRIG selector switch Trigger frequency ranges AC 20Hz...

Page 30: ...nector horizontal deflection in X Y mode Y POS II Controls vertical position knob of channel II display Inoperative in X Y mode POWER Turns scope on and off pushbutton LED LED indicates operating cond...

Page 31: ...1 3 5 6 7 8 9 14 15 16 17 2 4 10 11 12 13 20 21 22 23 24 25 26 27 28 30 31 34 36 37 38 33 32 18 29 19 35...

Reviews: