background image

Subject to change without notice

5

          I m p o r t a n t  h i n t s

Important hints

The operator is requested to read the following 
instructions and those of the mainframe HM8001-
2 carefully, to avoid any operating errors and 
mistakes and in order to become acquaint with 
the module.

After unpacking the blank module, check for any 
mechanical damage or loose parts inside. Should 
there be any transportation damage, inform the 
supplier immediately and do not put the module 
into operation. This plug-in module is primarily 
intended for use in conjunction with the mainfra-
me HM8001-2. When incorporating it into other 
systems, the module should only be operated with 
the specifi ed supply voltages.

Safety

This instrument has been designed and tested in 
accor dance with IEC Publication 1010-1, Safety 
require ments for electrical equipment for measu-
rement, control, and laboratory use. It corres-
ponds as well to the CENELEC regulations EN 
61010-1. All case and chassis parts are connected 
to the safety earth conductor. Corresponding to 
Safety Class 1 regulations (three-conductor AC 
power cable). Without an isolating transformer, 
the instrument's power cable must be plugged 
into an approved three-contact electrical outlet, 
which meets International Electrotechnical Com-
mission (IEC) safety standards.

Warning!
Any interruption of the protective conductor 
inside or outside the instrument or discon-
nection of the protective earth terminal is 
likely  to  render  the  instrument  dangerous. 
Intentional interruption is prohibited.

The istrument must be disconnected and secured 
against unintentional operation if there is any sug-
gestion that safe operation is not possible. 

This may occur:
–  if the instrument shows visible damage,
–  if the instrument has loose parts.
–  if the instrument does not function,
–  after long storage under unfavourable circ-

umstances (e.g. outdoors or in moist environ-
ments),

–  after excessive transportation stress (e.g. in 

poor packaging).

When removing or replacing the metal case, the 
instrument  must  be  completely  disconnected 
from the mains supply. If any measurement or 
calibration procedures are necessary on the 
opened-up instrument, these must only be carried 
out by qualifi ed personnel acquainted with the 
danger involved.

Symbols marked on equipment

 

ATTENTION  refer to manual.

 

DANGER!  High voltage.

 

Protective ground (earth) terminal.

Operating conditions

The ambient temperature range during operation 
should be b5°C and +40 °C and should 
not exceed –20 °C or +70 °C during transport or 
storage. The operational position is optional, how-
ever, the ventilation holes on the HM8001-2 and on 
the plug-in modules must not be obstructed.

Warranty and Repair

HAMEG instruments are subjected to a strict qua-
lity control. Prior to leaving the factory, each ins-
trument is burnt-in for 10 hours. By intermittent 
operation during this period almost all defects are 
detected. Following the burn-in, each instrument 
is tested for function and quality, the specifi cations 
are checked in all operating modes; the test gear 
is calibrated to national standards.

The warranty standards applicable are those of 
the country in which the instrument was sold. 
Reclamations should be directed to the dealer.

Only valid in EU countries

In order to speed reclamations customers in EU 
countries may also contact HAMEG directly. Also, 
after the warranty expired, the HAMEG service will 
be at your disposal for any repairs.

Summary of Contents for HM800

Page 1: ...B l a n k M o d u l e H M 8 0 0 Manual English...

Page 2: ...ions must have a double screening For IEEE bus purposes the double screened cable HZ72 from HAMEG is suitable 2 Signal cables Basically test leads for signal interconnection between test point and ins...

Page 3: ...rating conditions 5 Warranty and Repair 5 Maintenance 6 Operation of the blank module 6 Interface board for HM800 7 HM8001 2 Interface Speci cation 8 This will not cause damage or put the instrument o...

Page 4: ...e rails for mounting circuit boards at 4 different levels Plastic front panel for easy processing Power is supplied by the mainframe HM8001 2 Available supply voltages load capability see datasheet of...

Page 5: ...if there is any sug gestion that safe operation is not possible This may occur if the instrument shows visible damage if the instrument has loose parts if the instrument does not function after long s...

Page 6: ...hen reclosing the module care should be taken that the guides engage correctly with the front chassis Operation of the blank module Provided that all hints given in the operating instructions of the m...

Page 7: ...2 Without any resitance the output voltage amounts 5 2V The following table shows the resistance values for some common voltages VA 3 4 of CN401 R401 VB 5 6 of CN401 R402 5 2V No resistor 10V 3 92k 1...

Page 8: ...30 10 32 11 33 12 34 13 35 14 36 15 37 16 38 Adj 17 39 18 40 19 41 20 42 21 43 22 44 5V 4 26 BNC optional 230 V 115 V 8V 1A 5V 1A 5 20V 0 5A 7 VAC 18 VAC 7 VAC 18 VAC 8 VAC 0 4 A HM8001 2 Interface S...

Page 9: ...19 2V 11 33 Between Pin 11 12 7VAC 0 8Amax Potential 2 7 0V 6 6V 12 34 Between Pin 10 12 25VAC 0 8Amax Potential 2 27 7V 13 35 Supply of U2 U2a not used connect PIN 11 13 U3 14 36 5 20V regulated Ref...

Page 10: ...Subject to change without notice 10 Notes...

Page 11: ...Subject to change without notice 11 N o t e s...

Page 12: ...Mainhausen A Rohde Schwarz Company Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr DE 071040 QM sales hameg com authorized dealer Oscilloscopes Spectrum Analyz...