STOP
Do not cover either the holes of the case nor the
cooling fi ns.
Nominal specs are valid after a warm-up period of min. 20 min.
in the interval of +15 to +30 degrees C. Values without a tolerance
are typical of an average production instrument.
Warranty and Repair
HAMEG instruments are subject to a strict quality control. All
instruments are burned in for 10 hrs prior to shipment. By in-
termittent operation almost all early failures are detected. After
burn-in a thorough test of all functions and of quality is run, all
specifi cations and operating modes are checked.
In case of reclamations during the two years warranty period
please contact the dealer from whom you purchased your
HAMEG instrument. Customers from the Federal Republic of
Germany may directly contact HAMEG for warranty processing
in order to speed up the procedure.
The proceeding of repairs during the warranty period is subject
to our terms of warranty which are available on our website
http://www.hameg.com. Even after expiry of the warranty period
please do not hesitate to contact our HAMEG customer service
for repairs and spare parts.
Return Material Authorization (RMA):
Before sending back your instrument to HAMEG do apply
for a RMA number either by fax or on the Internet: http://
www.hameg.de.
If you do not have suitable packaging for the instrument
on hand please contact the HAMAG sales department
(Tel.: +49 (0) 6182/800 300, E-mail: vertrieb@hameg.de)
to order an empty original cardboard box.
Maintenance
The instrument does not require any maintenance. Dirt may
be removed by a soft moist cloth, if necessary adding a mild
detergent. (Water and 1 %.) Grease may be removed with ben-
zine (petrol ether). Displays and windows may only be cleaned
with a moist cloth.
Do not use alcohol, solvents or paste. Under no
circumstances any fl uid should be allowed to get into
the instrument. If other cleaning fl uids are used da-
mage to the lacquered or plastic surfaces is possible.
Line voltage selector
The instrument is destined for operation on 115 or 230 V mains,
50/60 Hz. The proper line voltage is selected with the line voltage
selector. It is necessary to change the fuse observing the proper
values printed on the back panel.
Please note:
After changing the main volage, the line fuse has
to be changed. Otherwise the instrument may be
destroyed.
Change of fuse
The mains fuse is accessible on the back panel. A change of
the fuse is only allowed after the instrument was disconnected
from the line and the power cord removed. Fuse holder and
power cord must not show any sign of damage. Use a screw
driver to loosen the fuse holder screw counterclockwise while
pressing the top of the fuse holder down. The top holding the
fuse will then come off. Exchange the defective fuse against a
correct new one.
It is forbidden to repair defective fuses or to bridge
them by any means. Any damage caused this way
will void the warranty.
Type of fuse:
5 x 20 mm; 250V~, C;
IEC 127/III; DIN 41662
(DIN 41571/3).
Value
115 V: 0.5 A slow blow
230 V: 1.0 A slow blow