4. Fitting of the HO720

Precondition! 
The installation of the interface HO720 is meaningful only if, 
as described under point 3 (Firmware), the oscilloscope has 
fi rmware version 03.000-yy.yyy or higher. Otherwise the in-
terface is not recognized by the oscilloscope. With spectrum 
analyzers the interface will be recognized by each fi rmware 
version.

STOP

 Safety!

 

The following procedures must only be carried out 
on condition that the mains (line) power cable is not 
connected to the oscilloscope and/or the spectrum 
analyzer and no connection is made at the mea-
surement inputs.   

STOP

 Attention!

 

To avoid damage of the interface during removing 
and fi tting by electrostatic discharge, please link a 
metal part of the oscilloscope / spectrum analyzer 
to equalise potentials between oscilloscope / spec-
trum analyzer and your body. Maintain this connec-
tion during the fi tting/removing! 

Only touch the interface at its mounting panel!

4.1 Removing the existing interface

4.1.1

 Remove both fastening screws. 

4.1.2

 Pull out the interface.

4.2 Fitting the interface HO720

4.2.1

 Insert the interface HO720 

in the opening in such a way, that 
the PCB will be inserted in the 
guides – visible on both sides – and push it in completely.  

4.2.2

 Fit the interface with the fastening screws previously (item 

4.1.1) removed. 

          F i t t i n g  o f  t h e  H O 7 2 0

Summary of Contents for HO720

Page 1: ...D U A L I N T E R F A C E H O 7 2 0 Installation guidelines English Issued September 2006...

Page 2: ...result from such conditions in individual cases HAMEG Instruments GmbH KONFORMIT TSERKL RUNG DECLARATION OF CONFORMITY DECLARATION DE CONFORMITE DECLARACI N DE CONFORMIDAD Hersteller Manufacturer Fab...

Page 3: ...garding the CE marking 2 1 Safety hints 4 2 Interface description 4 2 1 USB 2 2 RS 232 3 Firmware 4 4 Interface tting instruction 5 4 1 Removing the existing interface 4 2 Fitting the interface HO720...

Page 4: ...via a USB hub a USB cable is required equipped with Type B male connector one end and a Type A male connector at the other 2 2 RS 232 The RS 232 connector and its speci cation comply with the interfac...

Page 5: ...the mea surement inputs STOP Attention To avoid damage of the interface during removing and tting by electrostatic discharge please link a metal part of the oscilloscope spectrum analyzer to equalise...

Page 6: ...isplayed It enables RS 232 interface parameter settings that must correspond with the RS 232 settings in a terminal program or the HMLab settings I n t e r f a c e s e l e c t i o n 5 2 Spectrum analy...

Page 7: ...rdware Wizard is displayed Only in this case the USB Driver must be installed The USB Driver is located in the directory HO720_D2xx_Driver subdirectory Driver 6 1 Installation under Windows XP 6 1 1 P...

Page 8: ...3 4 After the selected path is displayed click Next 6 1 4 Thereafter the Hardware Installation window is displa yed showing a warning to continue the installation procedure As this warning is irrelev...

Page 9: ...Software Firmware below the oscilloscope picture HMLab zip Programming commands The programming commands list will be provided in the Internet under www hameg com 7 2 Spectrum analyzers The interface...

Page 10: ...riestra e 6 A Rohde Schwarz Company D 63533 Mainhausen registered trademark Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de authorized...

Reviews: