51
Calibrate the scanner
The bar code scanning device in the SP400X is a model 5180 imager that was manufactured by Honeywell Imaging and Mobility.
To understand the imager better or to modify settings, refer to the 5X80 engine Series 5X10/5X80 User’s Guide that can be downloaded at
www.honeywell.com/aidc
.
The 5X10/5X80 User’s Guide contains configuration bar codes that can be used to customize the desired allowable bar codes and to change
options and settings on the imager.
Reset the standard product defaults
If you are not sure what programming options are in your imager, or if you have changed some options and want the factory settings resto-
red, scan the Standard Product Default Settings bar code, below.
fIgURE 10-10
Standard product default settings bar code.
NOTE:
The light bar will briefly go out to indicate that the scan was completed successfully.
Update firmware through USB
1. Connect the SP400X to a computer with a USB cable.
2. open the Handheld SP400X Series Configuration Software. See ”
DOWNlOAD AND INsTAll sOfTWARE.
”
3. Select the Update tab.
There are four types of firmware update files:
• Firmware: the main program running on the SP400X
• FPGA: the processor configuration
• Imager: the program running on the bar code scanner module
• Communication: the program running on the wireless communication module
4. In the file boxes, enter the names of the update files. you can use the Browse button to navigate to the update files.
NOTE:
All of the update file names must have an “.hhp” extension.
5 . Press the Write to Device button to begin the update process.
6. If you are doing a firmware or fPGA update, a dialog box instructing you to restart the SP400X appears:
a. Click OK to dismiss the dialog box.
b. Press the green Power button on the SP400X and hold for three seconds to turn the device off.
c. After the SP400X has turned off, restart it by pressing the Power button for one second.
d. When the connection status indicates that the device is connected again, restart the update process by clicking the Write To Device button.
7. A dialog box announcing that the update has been successful or has failed appears. If the update failed, try performing the update again before
contacting support personnel.
SP400X Series - 10. Advanced use