6 - 24
Quick Check® SV Series User’s Guide
This re-initialization procedure recalls all of the current setup parameters from
the flash.
~SY
Software Sync for communication output modes #1 and #2 (~HO1, ~HO2);
trigger a sync response from the verifier, and enable the verifier
~Sh##
Set the high target reflectance value for calibration.
## - The high target reflectance value (70 to 99).
This sets a new high reflectance target value for a calibration symbol.
To use the target values encoded in the calibration symbol set this value to 00.
~Sl##
Set the low target reflectance value for calibration.
## - The low target reflectance value (1 to 20).
This sets a new low reflectance target value for a calibration symbol.
To use the target values encoded in the calibration symbol set this value to 00.
Output Interface Modes Descriptions
Output interface modes are used to set the SV Series for particular ways to
activate output ports. The mode is set via the ~LV## command. The following
list describes the standard output interface modes available in SV firmware
versions x270 and higher. Each description includes the commands for
programming the particular parameters that can activate the proper ports for that
mode. Custom modes are available – contact Hand Held Products Technical
Support.
After programming the SV unit for the desired port activation parameters, it is
recommended that you review them. When using ScanView, this can be done
via the Transmit SV Commands Window by sending the ~HT command to review
all scanner settings and sending the ~PT8 command to review all quality
parameter failure threshold settings.
Mode 00 (~LV00)
This mode does not activate output ports. Only the serial communications port
is active.