6 - 42
Quick Check® SV Series User’s Guide
~PB816xxx100
This command sets the minimum percent of scans on a code which calculate
good quiet zones to determine an acceptable quiet zone analysis. The field xxx
is the minimum passing threshold. Example: if xxx = 030, a minimum of 30% of
all fully decoded scans on a code must calculate a good quiet zone, or a failure
condition is set for this parameter.
~PB806xxx000
This command sets the threshold for the ANSI Defects calculation on a code to
set a failure condition. The field xxx is the passing threshold for the calculation.
For example: if xxx = 025, a Defects analysis of 26% or higher will cause a failure
condition for this parameter.
~PB802xxx100
This command sets the threshold for the ANSI Decodability calculation on a code
to set a failure condition. The field xxx is the passing threshold for the
calculation. For example: if xxx = 037, a Decodability analysis of 36% or lower
will cause a failure condition for this parameter.
~PB804xxx100
This command sets the threshold for the ANSI Symbol Contrast calculation on a
code to set a failure condition. The field xxx is the passing threshold for the
calculation. For example: if xxx = 020, a Symbol Contrast analysis of 19% or
lower will cause a failure condition for this parameter.
~PB815xxx000
This command sets the threshold for the ratio calculation on a code to set a
failure condition. The field xxx is the passing threshold for the calculation. For
example: if xxx = 025, a ratio analysis of 2.4 or lower will cause a failure condition
for this parameter.
~PB811xxxyyy
This command sets the threshold for acceptable X dimension ranges. The value
for xxx is the narrowest acceptable X dimension. The value yyy is the widest
acceptable X dimension. Units are in .001 inch (mils). For example:
~PB811010020 will cause a failure for an X dimension less than 10 mils or
greater than 20 mils.
~LN##
This command sets the minimum number of codes to be read during a sync
period. ## = the number of codes. For example: ~LN02 causes a No Read
condition to be set if less than 2 bar codes are fully decoded during a sync
interval. See the ~LZ## command below for priority.
If the value ## = 00 (and ~LZ00 is also set), a sync input is disregarded, therefore
a No Read condition is disabled.
~LZ##
This command sets the exact number of codes to be read during a sync period.
## = the number of codes. For example: ~LZ02 causes a No Read condition to
be set if exactly 2 bar codes are not fully decoded during a sync interval.