To the height of the lowest point one has to add successively:

•  47.24 inch / 120.00 cm  from the seat of the person to the swivel hook

•  3.35 inch / 8.50 cm  from the swivel hook to the middle of the arm

•  11.81 inch / 30.00 cm  for the lowest arm position vis-à-vis the horizontal 

position

•  0.39 inch / 1.00 cm  from the horizontal position to the screws.
One also has to take into account that in this way the person only just touches 

the bath. If you want to reach a sitting position and if you want to be able to 

fasten or loosen the sling straps, you have to deduct a further 7.87 inch / 20.00 

cm .
example of a transfer wheelchair-hospital bed.  The seat of the wheelchair is 

19.69 inch / 50.00 cm  high, the bed is 31.5 inch / 80.00 cm  high. That means 

the wheelchair is the determining point. The result is:  19.69 inch / 50.00 cm   

(wheelchair) + 47.24 inch / 120.00 cm  + 3.35 inch / 8.50 cm  + 11.81 inch / 

30.00 cm  - 0.39 inch / 1.00 cm  - 7.87 inch / 20.00 cm  = 73.82 inch / 187.50 

cm . The final positioning should be at 72.83 inch / 185.00 cm  height to the 

top bolt holes. 

a

19.69 inch / 50.00 cm  - height of the wheelchair (may vary)

B

7.87 inch / 20.00 cm  - for attaching/detaching the sling loops

c

47.24 inch / 120.00 cm  - from the seat of the person to the swivel hook

d

3.35 inch / 8.50 cm  - from the swivel hook to the middle of the arm

e

11.81 inch / 30.00 cm  - for the lowest arm position vis-à-vis the horizontal position

f

0.39 inch / 1.00 cm  - from the horizontal position to the screws

You can folloW thIs rule: heIght of the  loWest PoInt 

+ 53.15 Inch / 135.00 cm 

 (= uPPer fIxIng holes Wall Plate)

1.2.  minimum strength of the wall

a

51.18 inch / 1300.00 mm 

B

1450 n

c

4000 n

d

1700 n

C

E

B

D

A

F

A

B

A

B

C

C

C

C

D

D

Summary of Contents for Wall lift

Page 1: ...ers Wall lift Installation Table of contents 1 On the wall p 2 1 1 Installation instructions p 2 1 2 Minimum strength of the wall p 3 2 On a pole p 4 2 1 Installation instructions p 4 2 2 Minimum strength of the ceiling and or wall p 4 ...

Page 2: ... in the direction of the highest position Distance from wall to swivel hook Position Lowest inch cm Horizontal inch cm Highest inch cm 1 36 54 92 80 37 4 95 00 33 46 85 00 2 37 01 94 00 39 37 100 00 35 24 89 50 3 38 86 98 70 41 34 105 00 36 97 93 90 4 40 67 103 30 43 31 110 00 38 7 98 30 5 42 52 108 00 45 28 115 00 40 43 102 70 6 44 33 112 60 47 24 120 00 42 17 107 10 7 46 18 117 30 49 21 125 00 4...

Page 3: ...igh the bed is 31 5 inch 80 00 cm high That means the wheelchair is the determining point The result is 19 69 inch 50 00 cm wheelchair 47 24 inch 120 00 cm 3 35 inch 8 50 cm 11 81 inch 30 00 cm 0 39 inch 1 00 cm 7 87 inch 20 00 cm 73 82 inch 187 50 cm The final positioning should be at 72 83 inch 185 00 cm height to the top bolt holes A 19 69 inch 50 00 cm height of the wheelchair may vary B 7 87 ...

Page 4: ... and the wall at the top A B C D E F G H I J K L M N O P Q R S T U V P inch mm inch mm inch mm inch mm A 2 76 70 00 G 0 87 22 00 M 0 16 4 00 S 2 36 60 00 B 1 57 40 00 H 1 97 50 00 N 3 62 92 00 T 6 69 170 00 C 6 69 170 00 I 0 79 20 00 O 2 62 66 50 U 0 39 10 00 D 2 95 75 00 J 25 59 650 00 P 20 35 517 00 V 0 79 20 00 E 1 97 50 00 K 4 72 120 00 Q 2 76 70 00 F 0 24 6 00 L 1 97 50 00 R 1 57 40 00 2 2 Mi...