10
Molex testing link
5. Testing
The circuits should be tested as follows:
• Use the electrical diagram to see which circuit you need to test
• Disconnect the Handicare 2000 from the power supply
• Connect the molex testing link to the pcb
• Ensure looms are removed from pcb before carrying out
continuity testing
• Set the multimeter to the “beep” or resistance or voltage test
mode
• Check that you are testing between the correct connector pins/
numbers.
• No “beep” or resistance means that the circuit being tested is
not complete.
• Voltages can be find on the electrical scheme
If there is a short circuit with the frame, hold one of the measuring pins
against the unit’s ground pin of the seat and push the other measuring
pins into the top of the molex testing link.