Serial number

Maintenance

The lift must undergo thorough inspection at least once per year. Inspection must be performed by authorized personnel 
and in accordance with Handicare’s service manual.
Repairs and maintenance may only be done by authorized personnel using original spare parts.

 

Used batteries are to be left at the nearest recycling station. Used batteries can also be returned to 

 

Handicare or a Handicare dealer for recycling.

+50 ºC

-10 ºC

Cleaning/disinfection

If necessary, clean the lift with warm water or a soap solution and check that the castors are free from dirt and hair. Do 
not use cleaning agents containing phenol or chlorine, as this could damage the aluminium and the plastic materials. 
If disinfection is needed, 70 % ethanol, 45 % isopropanol or similar should be used. Disinfect the parts that come in 
contact with a user before the lift is used for the next user.

Storage and transportation

If the lift is not to be used for some time or e.g., during transport, we recommend that the emergency stop button be 
pressed in. The lift should be transported and stored in -10 ° C to + 50 ° C and in normal humidity, 30% -75 %. The air 
pressure should be between 700 and 1060 hPa. Let the lift reach room temperature before the batteries are charged or 
the lift is used.

Service agreements

Handicare offers service agreements for maintenance and regular testing of your lift unit. Contact your local Handicare 
representative.

106 kPa

70 kPa

75%

30%

1060 hPa

700 hPa

+50 °C

-10 °C

75%

30%

14

I N S T R U C T I O N S   F O R   U S E System

RoMedic

TM

Summary of Contents for SystemRoMedic EvaDrive

Page 1: ...oMedic TM Instructions for use English Bruksanvisning Svenska Brukermanual Norsk Brugsvejledning Dansk K ytt ohje Suomi Gebrauchsanweisung Deutsch Handleiding Nederlands Manuel utilisateur Fran ais Ma...

Page 2: ...driving function in all directions Check that the accessory EvaDrive is properly mounted on the lift Always read the user manuals for all assistive devices used during a transfer Keep the user manual...

Page 3: ...0 mm Installation of the handlebar The handlebar must be installed on the mast at a height of at least 33 4 inch from the floor Measure from the lowest point of the handlebar The handlebar is secured...

Page 4: ...press the down button and run the boom all the way down Test base width adjustment Activate the system press the button for base width adjustment open the undercarriage to maximum width and then pres...

Page 5: ...tions are made by authorized personnel Never attempt to lift the mobile lift unit by the handlebar Ensure that there are no obstacles or people in the way of the lift Never stand on the lift during op...

Page 6: ...unction is described in the respective section in the user manual or in the user manual for the respective accessory Below is a brief description of the functions that are available for caregivers and...

Page 7: ...level Solid light indicates battery capacity remaining Blinking light indicates error A service technician can determine the source of the error Power button Press lightly to activate EvaDrive Wait ab...

Page 8: ...eration of the lift is not possible until EvaDrive has completed automatic calibration If the handlebar is hit violently or subjected to impact the sensors neutral values can be affected so much that...

Page 9: ...top blinking All or parts of the battery scale blink quickly Something is wrong with the lift s electrical system Restart EvaDrive If there is still a problem contact Handicare Red light battery scale...

Page 10: ...ned see Fig 2 The yellow LED light on the control panel indicates that the wheels are in the turned position Grasp the handlebar and hold in the safety switch then turn the lift in the desired directi...

Page 11: ...rive cover Emergency stop switches off all of the lift s electrical functions The battery scale continues to light continuously Resetting Turn the emergency stop button in the direction of the arrows...

Page 12: ...ge the aluminium and the plastic materials If disinfection is needed 70 ethanol 45 isopropanol or similar should be used Disinfect the parts that come in contact with a user before the lift is used fo...

Page 13: ...ght Eva450Drive 127 87 lbs 58 kg Eva600Drive 141 1 lbs 64 kg Weight heaviest part Eva450Drive 97 lbs 44 kg Eva600Drive 110 2 lbs 50 kg Turning diameter Eva450Drive 54 inch 1340 mm Eva600Drive 61 8 inc...

Page 14: ...All measurement are in mm inch Tolerance 5 mm 0 2 When the handle is placed 1000 mm from floor Eva450Drive 25 0 9 427 16 8 1307 51 4 1450 57 1 240 9 4 900 1100 35 4 43 3 124 I N S T R U C T I O N S F...

Page 15: ...n mm inch Tolerance 5 mm 0 2 Eva600Drive When the handle is placed 1000 mm from floor 806 31 7 395 15 5 25 0 9 432 17 1408 55 4 1566 61 6 240 9 4 900 1100 35 4 43 3 125 I N S T R U C T I O N S F O R U...

Page 16: ...es RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class A The EvaDri...

Page 17: ...e 2 kV for common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000...

Page 18: ...occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnet...

Page 19: ...ncy of transmitter m 150 kHz to 80 MHz P d 2 1 80 MHz to 800 MHz P d 2 1 800 MHz to 2 5 GHz P d 3 2 0 01 0 12 0 12 0 24 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters ra...

Page 20: ...oMedic offers improvement of both work environment and quality of care and at the same time achieves significant cost savings Always make sure that you have the correct version of the manual The most...

Reviews: