background image

Medical Device Class I. The product complies with the requirements of the Medical Devices Directive 93/42/EEC.

Maximum speed (normal) 

Forward: 4.3 mph / 7 km/h, Backwards: 2.1 mph / 3,5 km/h

Maximum speed (low speed ) 

2.1 mph / 3,5 km/h

Batteries

24V, 17,6 Ah Li-Ion 

Charger

Modiary MDA10129402000. Output: DC 29, 40 V, 2.0 A

AC power input cable

HD3VV-F 3G0,75mm². TCA090812EA. Length 59 inch, 1500 mm

Motor cover

Flame-resistant ABS plastic

Intermittent operation

Op 10/90, active op. max. 2 min. Out of a time of 100, active time must be less than 
10, though not more than 2 min.

Nav motor

24V DC Mono Shaft, 180 W

Material

Aluminium/steel

Emergency lowering

Manual

Castors

8 inch, 203 mm

IP class

IP X4*

Weight

Eva450Drive:                         127.87 lbs, 58 kg
Eva600Drive:                         141.1 lbs, 64 kg

Weight (heaviest part)

Eva450Drive:                         97 lbs, 44 kg
Eva600Drive:                         110.2 lbs, 50 kg

Turning diameter

Eva450Drive:                         54 inch/ 1340 mm 
Eva600Drive:                         61.8 inch/ 1570 mm

Expected service life

10 years

Pressure control panel

4 N

Clearance

1.1 inch, 28 mm

*IP X4: Splashing of water, Water splashing against the enclosure from any direction shall have no harmful effect.  

(Test duration: 5 minutes Water volume: 10 litres per minute Pressure: 50–150 kPa)

Technical information

This equipment/system is intended for use by healthcare professionals only. This equipment/system may cause radio 
interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such 
as reorienting or relocating the equipment/system or shielding the location.

EMC

15

I N S T R U C T I O N S   F O R   U S E

System

RoMedic

TM

English

Summary of Contents for SystemRoMedic EvaDrive

Page 1: ...oMedic TM Instructions for use English Bruksanvisning Svenska Brukermanual Norsk Brugsvejledning Dansk K ytt ohje Suomi Gebrauchsanweisung Deutsch Handleiding Nederlands Manuel utilisateur Fran ais Ma...

Page 2: ...driving function in all directions Check that the accessory EvaDrive is properly mounted on the lift Always read the user manuals for all assistive devices used during a transfer Keep the user manual...

Page 3: ...0 mm Installation of the handlebar The handlebar must be installed on the mast at a height of at least 33 4 inch from the floor Measure from the lowest point of the handlebar The handlebar is secured...

Page 4: ...press the down button and run the boom all the way down Test base width adjustment Activate the system press the button for base width adjustment open the undercarriage to maximum width and then pres...

Page 5: ...tions are made by authorized personnel Never attempt to lift the mobile lift unit by the handlebar Ensure that there are no obstacles or people in the way of the lift Never stand on the lift during op...

Page 6: ...unction is described in the respective section in the user manual or in the user manual for the respective accessory Below is a brief description of the functions that are available for caregivers and...

Page 7: ...level Solid light indicates battery capacity remaining Blinking light indicates error A service technician can determine the source of the error Power button Press lightly to activate EvaDrive Wait ab...

Page 8: ...eration of the lift is not possible until EvaDrive has completed automatic calibration If the handlebar is hit violently or subjected to impact the sensors neutral values can be affected so much that...

Page 9: ...top blinking All or parts of the battery scale blink quickly Something is wrong with the lift s electrical system Restart EvaDrive If there is still a problem contact Handicare Red light battery scale...

Page 10: ...ned see Fig 2 The yellow LED light on the control panel indicates that the wheels are in the turned position Grasp the handlebar and hold in the safety switch then turn the lift in the desired directi...

Page 11: ...rive cover Emergency stop switches off all of the lift s electrical functions The battery scale continues to light continuously Resetting Turn the emergency stop button in the direction of the arrows...

Page 12: ...ge the aluminium and the plastic materials If disinfection is needed 70 ethanol 45 isopropanol or similar should be used Disinfect the parts that come in contact with a user before the lift is used fo...

Page 13: ...ght Eva450Drive 127 87 lbs 58 kg Eva600Drive 141 1 lbs 64 kg Weight heaviest part Eva450Drive 97 lbs 44 kg Eva600Drive 110 2 lbs 50 kg Turning diameter Eva450Drive 54 inch 1340 mm Eva600Drive 61 8 inc...

Page 14: ...All measurement are in mm inch Tolerance 5 mm 0 2 When the handle is placed 1000 mm from floor Eva450Drive 25 0 9 427 16 8 1307 51 4 1450 57 1 240 9 4 900 1100 35 4 43 3 124 I N S T R U C T I O N S F...

Page 15: ...n mm inch Tolerance 5 mm 0 2 Eva600Drive When the handle is placed 1000 mm from floor 806 31 7 395 15 5 25 0 9 432 17 1408 55 4 1566 61 6 240 9 4 900 1100 35 4 43 3 125 I N S T R U C T I O N S F O R U...

Page 16: ...es RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class A The EvaDri...

Page 17: ...e 2 kV for common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000...

Page 18: ...occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnet...

Page 19: ...ncy of transmitter m 150 kHz to 80 MHz P d 2 1 80 MHz to 800 MHz P d 2 1 800 MHz to 2 5 GHz P d 3 2 0 01 0 12 0 12 0 24 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters ra...

Page 20: ...oMedic offers improvement of both work environment and quality of care and at the same time achieves significant cost savings Always make sure that you have the correct version of the manual The most...

Reviews: