18
S E R V I C E M A N U A L
Battery and cable installations
- Check battery and electrical cable
.
- Verify that all cables are properly inserted.
Controlbox
-Verify that all functions on the control panel are working
Handcontrol
-Verify that all functions on the handset are working.
Emergency stop
-Verify that the lift does not operate when emergency stop is pressed in.
Emergency lowering
-Verify the function of emergency lowering.
Charging
-Verify that charging is operating. Yellow lamp on the controllbox should light up when charging.
Product label
- Check the serial number so that the label is there and readable. Note the serial number on the protocol.
Test of safe working load
Mast safe working load
-Run the lift with safe working load, all the way up and down, listen for peculiar noises and vibrations.
Base safe working load
-Widen the base with safe working load, listen for peculiar noises and vibrations.
Accessories
-Check the accessories to the lift.
Documentation
- Check that the manual is available and up-to-date.
To reset the service counters press both lift buttons on the hand control at the same time for 5 seconds.
An audio signal will indicate that the timer has been reset.