3. After treatment

Understanding the symbols

www.hrhealthcare.co.uk/flowox

After the treatment turn off the 

device by pressing the ‘On/Off’ 

button.

E2 Major Air Leakage

This icon may appear once you turn on the device, which 

is safe. After a couple of minutes if this symbol and the E2 

symbol light up together, you may have higher air leakage 

than normal.

Roll down the seal onto the outside 

of the pressure chamber.

E3 Low Air Volume 

Not enough air to function.  

(Low air volume).

Press the blue release valve button to remove the air. To remove the air 

quicker, press down on the inner padding. Remember to roll back the seal off 

the top of the pressure chamber to ensure it doesn’t lose its stretch  (see 2d). 

Face Icon

A smiley face icon may 

appear if the device has 

been used for longer than 

2 hours. 

Time Card

You will get a time card of 

200 hours use, this symbol is 

a warning that you have 25 

hours of use left.

Solution:

 Check for 

obstruction, e.g. kinked 

hose. Restart the device by 

pressing the ‘On/Off’ button.

Solution:

 None-required, 

device is safe to use e.g. in 

the event you want to treat 

both legs.

Solution:

 Contact your 

Clinical Support Specialist 

from H&R Healthcare.

Solution: 

Turn off, remove foot, make sure the padding 

is fixed properly and that the seal is properly placed 

around the top of the pressure chamber. Follow steps 

2a to 2e.