18
OMNICYCLE
®
USER MANUAL
COPYRIGHT 2007 - 2013, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
Full Program Menu
The parameters of the exercising programs and systems settings are set out within the following menu structure:
1
st
Level
2
nd
Level
Possible Settings, Functions
Factory/Default Settings
NEURO
Duration
1 to 60 minutes
15 minutes
Forward/Backward Auto
On/Off
Off
Biofeedback Symmetry
bar/Symmetry-road, Passive, Off
Symmetry bar
Motor power
2 to 22 Nm
15 Nm
Spastic detector
Fine / Medium / Rough
Medium
Spastic autom.
Forward/Backward/Change
Forward
Speed adaption
On/Off
Off
Pulse demand
60 to 160 bpm
150 bpm
ORTHO
Duration
1 to 60 minutes
15 minutes
Biofeedback Symmetry
bar/Symmetry-road, Passive, Off
Symmetry bar
Motor power
2 to 22 Nm
15 Nm
Muscle support
Off/On
On
Pulse demand
60 to 160 bpm
150 bpm
CARDIO
Duration
1 to 60 minutes
15 minutes
Biofeedback Symmetry
bars/road,
Assisted, Off
Symmetry bar
Motor power
2 to 22 Nm
15 Nm
Pulse demand
60 to 160 bpm
120 bpm
SYSTEM
Sensitivity mike
1 to 10
2
Language
Multiple
Languages
English
Contrast
1 to 40
15
Description of Omnicycle
®
Functions
The first menu level comprises the following items:
•
Neuro
•
Ortho
•
Cardio
•
System (Technical)
Neuro
(basic setting): Exercise programs for patients with neurological conditions, e.g. hemiplegia, paraplegia, MS,
and Parkinson’s disease. The method of exercising is set as
Set resistance
. The resistance level set remains constant
whatever the pedaling speed. Higher pedaling speeds mean greater power output.
Ortho:
Exercising programs for orthopedic patients, e.g. following a knee or hip surgery. In this exercise mode,
the method of exercising is preset to
Set power
. The patient is given a set power level at which to exercise. The
Omnicycle
®
modulates the resistance automatically according to the pedaling speed so that the amount of power
required is always the same.