background image

Technical Support :+82-2-3452-8966 / techsupport@ihanil.com

21/23

6.2 Error Messages

6. Troubleshooting

Message

Cause

F RAM, SEEP ROM 

Check!

Occurs the data storage memory failure.

Contact Technical Support.

Over Temprature!

Occurs the temperature is above the allowable limit.

Contact Technical Support.

Doorlock Error

Occurs the lid sensor failure.

Contact Technical Support.

Thermal Sensor Error

Occurs the temperature sensor failure.

Contact Technical Support.

Hi Limit Temperature!

Occurs the current temperature is higher than temperature 

limit range.

Contact Technical Support.

Low Limit 

Temperature!

Occurs the current temperature is lower than temperature 

limit range.

Contact Technical Support.

Start RPM Error

Occurs when 100 RPM is not reached within 2 seconds after 

start.

Contact Technical Support.

Low RPM Error

Occurs when the setting rpm is more than + 2%.

Contact Technical Support.

Over RPM Error

Occurs when the setting rpm is more than - 2%.

Contact Technical Support.

Invert Fail

Occurs when an inverter failure.

Contact Technical Support.

Imbalance Error

Occurs wheh imbalnce is detected or imbalance sensor 

failure.

- Make sure the loaded blood bag adaptors opposite each 

  other on the rotor are the same weight/volume

- Contact Technical Support.

Summary of Contents for Component R12

Page 1: ...Operation Manual Component R12 Large Capacity Centrifuge Product Model Name Date of Purchase...

Page 2: ...018 Hanil Scientific Inc If you have any questions contact our Technical Support Center 82 2 3452 8966 techsupport ihanil com www ihanil com The appearance or specifications of the device is subject t...

Page 3: ...Using the Rotor 11 3 6 Loading the Blood Bags 12 4 Operation 13 4 1 Control Panel 13 4 2 Setting Speed 14 4 3 Setting Run Time 15 4 4 Setting Temperature 15 4 5 Setting Temperature Limit 15 4 6 Setti...

Page 4: ...le or explosive vapor can not be deviced 7 The balancing work of samples should done in advance before operation 8 To ensure safe use of the device do not expose the device to strong acids strong base...

Page 5: ...of a power cord Breakdown of circuit breaker fuse or safety device If you spill liquid on the equipment If the equipment has been damaged 1 2 Transport Storage 1 3 Safety label on the Device 1 4 Elect...

Page 6: ...3 2 Product Description 2 1 Structure Component R12 Operating manual T wrench Leveler tool Rotors Accessories on request 2 2 Delivery package Lid RPM measuring window Control panel Power switch key Em...

Page 7: ...Time 100 hr continuous Max Capacity Swing out rotor 6 x 2 400 mL Temperature range 10 to 40 ACC DEC steps 10 10 steps Program memory 100 Rotor Identification Automatic Imbalance cutoff tracking Yes N...

Page 8: ...ft may be bent due to the weight of the rotor Good ventailation For air circulation and safety maintain a 30cm clearance envelope around the centrifuge while the rotor is spinning Constant temperature...

Page 9: ...e If the device is not installed horizontally it may cause vibration noise or malfunction 2 Put the leveler tool on the device 2 Turn the leveling foot on the botton of the device to level the device...

Page 10: ...el attached to the device Only use a socket proviced Hanil Scientific Inc If the lid is not closed the device does not operate 1 You can choose a embedded socket or a movable socket 2 Switch the centr...

Page 11: ...82 2 3452 8966 techsupport ihanil com 11 23 This device dose not necessary to remove the rotor after use Do not disassemble the rotor by the user 3 Installation 3 5 Using the Rotor Do not disassemble...

Page 12: ...n Only use accessories provided by Hanil Scientific Inc 1 Load the blood bags into the adaptors with no folds at the top or corners 2 There are two holes in the each adaptor Load one blood bag into ea...

Page 13: ...ttons Enter a value using the numeric buttons View Setting Verifying the setting parameters during centrifugation Enter Setting acceleration step Program Save Saving a program Program Call Press the e...

Page 14: ...Press the Speed button once 2 Press the numeric buttons to set the desired RPM value Up to 4 800 rpm 3 Press the Enter button Setting RCF 1 Press the Speed button twice 2 Press the numeric buttons to...

Page 15: ...it manually 3 Press the Setup Enter button 4 3 Setting Run Time 1 Press the Temp button 2 Press the numeric buttons to set the desired temperature 10 40 3 Press the Setup Enter button 1 Press the Tem...

Page 16: ...n 4 Operation Saving 1 Set the RPM or RCF Run time Accel Decel rate Temperature 2 Press the Program Save button 3 Press the numeric buttons to set the desired program number Up to 0 99 100 4 Press the...

Page 17: ...3 During operation you can verify the setting values by pressing the View Setting button 4 Operation 4 8 Start Stop a Run 4 9 Verifying the Setting value Start a run 1 Check all parameters are correct...

Page 18: ...aminated use a mild cleaning fluid to clean 2 Do not use aggressive chemicals on the device such as alchol benzene acetone or phenol 3 Make sure do not scratch the surface of deivice when cleaning it...

Page 19: ...1 Wipe up the moisture in the chamber with a soft cloth 2 Leave the centrifuge lid open after centrifugation Remove the condensation water 1 There is a drain hose at the bottom of the device Regularly...

Page 20: ...Centrifugation cannot be started Check the lid is closed completely Lid cannot be opened Press the Lid open button Check the power connection Lid cannot be closed Remove the dirt on the door latch and...

Page 21: ...n temperature limit range Contact Technical Support Low Limit Temperature Occurs the current temperature is lower than temperature limit range Contact Technical Support Start RPM Error Occurs when 100...

Page 22: ...ou can activte the emergency open manually 1 Wait for rotor to stop before activating the emergency open 2 Insert the T wrench into the emergency open hole and press it until the lid is opened 6 3 Eme...

Page 23: ...L mm Max height for tube fit mm Max RPM rpm Max RCF xg B2400 TRB500 2 500 mL blood bag 2 12 73 x 125 209 2 4 500 6 491 TRB450 2 450 mL blood bag 2 12 62 x 100 TRB250 2 250 mL blood bag 2 12 38 5 x 84...

Page 24: ...16 Arayuk ro Gimpo REP of Korea 82 2 3452 8966 technicalsupport ihanil com www ihanil com Doc No OMP5PDEN1801...

Reviews: