10.
Trouble Shootings
Cut on the bowl seal
1.
Is there any cut on the bowl bottom gasket or any scratch on the gasket contact regions?
Please remind during bowl bottom disassembly.
Problem on feeding
1.
Is there any blockage on the feed nozzle?
2.
Is there any leakage on the three wings into the bottom of the bowl? When the tension on the
three wing spring, please replace it with new.
3.
Is there any problem on the feed pipe line?
Damages on the tension pulley bearing
1.
Was there any leakage on the tension pulley gasket?
If so, please replace the ball bearing together with gasket.
Damages on the ball bearing of the bearing pulley
1.
If there is damage on the rubber coupling, please check the same case with when damages on
the Rubber coupling.
2.
Is there any cut on the rubber sealing? Please replace the rubber sealing whenever disassemble
the pulley cap.