REAR PANEL CONNECTIONS

Terminals #1: Probe

• Connect the 

HI 7634-00

 probe by following the wires

colour indications.

Terminals #2: Power Supply

• Model 

BL983320-0:

 connect the 2 wires of a 12 Vdc

power adapter to the ter12 Vdc and GND.

• Model 

BL983320-1:

 connect a 3-wire power cable to the

terminals while paying attention to the correct earth (PE),
line (L) and neutral (N1 for 115 V or N2 for 230 V) contacts.

Terminals #3: Dosing Contact

• This contact drives the dosing system, accordingly to the

selected setpoint.

Note

: The setpoint has a typical hysteresis value comparable

to the meter accuracy.

Terminals #4: Not Used Contact
Overtime system: jumper (#5) and trimmer (#6)

• This system allows the user to set a maximum dosing

period, by adjusting the rear trimmer from 5 (min) to
approx. 30 (max) minutes.

• When the set time is exceeded, any dosing action stops,

the LED indicator on the front panel will blink Red and the
LCD will show the “TIMEOUT” warning message. To exit
the overtime condition, set the OFF/Auto/ON switch to
“OFF” position, and then to “Auto” again.

• For disabling the overtime feature, simply remove the

jumper on the rear panel.

Note:

 The overtime system works only if the OFF/Auto/ON

switch is in “Auto” position.

OPERATING THE METER

Before proceeding make sure that:
• the meter has been calibrated;
• the setpoint value has been properly adjusted;
• all rear panel wiring and selections are correct;
• the Auto/OFF/ON switch is in the desired position.
Install or immerse the probe in the solution to be monitored,
then press the “MEAS” key (if necessary).
The LCD will show the EC (µS/cm) value. The LED indicator
will light up Green when the meter is in measurement mode
and dosing is not active, while will light up Orange/Yellow for
signaling that a dosing action is in progress.

FUNCTIONAL DESCRIPTION

Front panel

Rear panel

1. Connections for EC probe

2. Power supply terminal:

• for 

BL983320-0 

model: 12 Vdc adapter

• for 

BL983320-1

 model: 115 Vac or 230 Vac option

3. This contact acts as a switch for driving the dosing system (e.g.

dosing pump)

4. Not used contact

5. Jumper for enabling (jumper in) or disabling (jumper

removed) the overtime control

6. Trimmer for overtime setting (typically from 5 to 30 minutes)

1. Liquid Crystal Display

2. Switch for selecting dosing mode:

• OFF = dosing disabled
• Auto = automatic dosage, depending on setpoint value
• ON = dosing always active

3. “MEAS” key to set the instrument to measurement mode

4. “SET” key to display and set the setpoint value

5. “SET” trimmer to adjust the setpoint value

6. “CAL” trimmer

7. 3-colour LED indicator:

• Green = meter in measurement mode
• Orange/Yellow = dosing in progress
• Red, blinking = indicates an alarm condition

OPERATIONS

CALIBRATION

To calibrate the meter, proceed as follows:
• ensure the meter is in measurement mode;
• immerse the probe in 

HI 7033 

calibration solution (84

µS/cm);

• shake briefly and wait for reading to stabilize;
• adjust the “CAL” trimmer to read “84.0 µS” on the LCD.

SETPOINT

Press the “SET” key: the display will show the default or
previously adjusted value, together with the “SET” indication.
Using a small screwdriver adjust the “SET” trimmer until the
desired setpoint value is displayed.
After 1 minute the meter automatically returns to the normal
mode; or press the “MEAS” key.

PROBE MAINTENANCE

To improve the probe performance and prolong its life, it is
recommended to clean it regularly.
• Immerse the tip of the probe in 

HI 7061

 Cleaning

Solution at least for one hour.

• If a more thorough cleaning is required, brush the metal

pins with very fine sandpaper.

• After cleaning, rinse the probe with tap water and

recalibrate the meter.

•  When not in use, clean the probe before storing it.

ISTBL3320R3           04/05

All external cables connected to the rear panel
should end with cable lugs.

A circuit breaker (rated 6A max.) must be
connected in close proximity to the equip-
ment, and in a position easy to reach by the
operator, for disconnection of the instrument
and of all the devices connected to the relays.

 

 

 

Summary of Contents for BL 983320-0

Page 1: ...Instruction Manual BL 983320 0 BL 983320 1 Panel Mounted EC Indicators Controllers ...

Page 2: ...l the instrument has been observed to function correctly Any defective item must be returned in its original packing Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct operation of the meter Please read it carefully before using the instrument If you need additional technical information do not hesitate to e mail us at t...

Page 3: ... terminal blocks on the rear panel The probe is easy to clean and requires little maintenance Other features include automatic temperature compensation of readings single point calibration overtime control system multi colour LED for indicating if the meter is in measure ment dosing alarm condition possibility to set Off Auto On switch dosing action mode Two models are available BL 983320 0 powere...

Page 4: ...ASSEMBLING VIEW ...

Page 5: ...dependingonsetpointvalue ON dosing always active 3 MEAS key to set the instrument to measurement mode 4 SET key to display and set the setpoint value 5 SET trimmer to adjust the setpoint value 6 CAL trimmer 7 3 colourLEDindicator Green meter in measurement mode Orange Yellow dosing in progress Red blinking indicates an alarm condition ...

Page 6: ...ct 5 Jumper for enabling jumper in or disabling jumper removed theovertimecontrol 6 Trimmerforovertimesetting typicallyfrom5to30minutes All external cables connected to the rear panel should end with cable lugs A circuit breaker rated 6A max must be connected in close proximity to the equip ment and in a position easy to reach by the operator for disconnection of the instrument and of all the devi...

Page 7: ...es When the set time is exceeded any dosing action stops the LED indicator on the front panel will blink Red and the LCD will show the TIMEOUT warning message To exit the overtime condition set the OFF Auto ON switch to OFF position and then to Auto again For disabling the overtime feature simply remove the jumper on the rear panel Note The overtime system works only if the OFF Auto ON switch is i...

Page 8: ...l screwdriver adjust the SET trimmer until the desired setpoint value is displayed After 1 minute the meter automatically returns to the normal mode or press the MEAS key PROBE MAINTENANCE To improve the probe performance and prolong its life it is recommended to clean it regularly Immerse the tip of the probe in HI 7061 Cleaning Solution at least for one hour If a more thorough cleaning is requir...

Page 9: ... ß 2 ºC Calibration Manual through CAL trimmer Dosing Contact Maximum 2A fuse protected 250 Vac 30 Vdc Contact close when measure setpoint Setpoint Adjustable from 0 to 199 9 µS cm Overtime Adjustable typically from 5 to approx 30 minutes Power Consumption 10 VA Installation Category II Power supply External fuse protected BL983320 0 12 Vdc BL983320 1 115 230 Vac 50 60Hz Dimensions 83 x 53 x 99 mm...

Page 10: ... mL HI 7061M Probe cleaning solution 230 mL bottle HI 7061L Probe cleaning solution 500 mL bottle HI 710005 12 Vdc power adapter US plug HI 710006 12 Vdc power adapter European plug HI 710012 12 Vdc power adapter Australian plug HI 710013 12 Vdc power adapter South African plug HI 710014 12 Vdc power adapter UK plug HI 731326 Calibration screwdriver 20 pcs HI 740146 Mounting brackets ...

Page 11: ...anship and materials when used for their intended purpose and maintained according to instructions Probes are warranted for six months This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed main tenance are not covered If service is required contact the dealer from whom you purchased the instrument If under warranty report th...

Page 12: ...CE DECLARATION OF CONFORMITY ...

Reviews: