Operational Guide
29
pH meter operation is configured by using the SETUP key with a pH probe
connected to the meter. The parameter-specific parameters will be seen
inserted into the menu. If Basic mode is configured “On”, the pH parameter list
will not be displayed. See Basic mode for a description of this operation before
choosing how to SETUP the meter.
Parameter
Description
Choices Default
Basic mode
Basic Mode
When “On”, a limited set of
parameters and calibration
buffers are available for
use.
Off or
On
Off
Available
Information
Visual indication of buffers
used, Probe Condition
and Response times are
determined and displayed
when calibration uses 7 and
4 and/or 10 pH buffers.
Off or
On
On
Buffers used, probe
conditions, and
response times are
not displayed.
First Custom
Buffer
When “On”, it permits the
user to enter a custom pH
buffer value to use during
probe calibration.
Off or
On
Off
Not Available
Second
Custom Buffer
When “On”, permits user to
enter a custom pH buffer
value to use during probe
calibration.
Off or
On
Off
Not Available
First
Calibration
Point
Allows the user to choose
how the first point in
calibration will be made.
Offset
or point
Offset
Not Available
(automatically uses
Offset).
Resolution
Allows the user to select
between 0.01 and 0.001 pH
resolution.
0.01 Or
0.001
pH
0.01 pH
Not Available
(automatically uses
0.01 pH resolution).
Set Out Of
Calibration
Range
When “On”, a measurement
outside the calibration
buffer window used will
trigger a warning message.
Off or
On
On
Not Available, No
Error messages
displayed.
pH Meter
Configuations